  "results": [
    {
      "status": "Terminated",
      "city": "Feldkirchen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67719",
      "recalling_firm": "Brainlab AG",
      "address_1": "Kapellenstrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France,  Germany, Japan, Netherlands, and the United Kingdom.",
      "recall_number": "Z-1513-2014",
      "product_description": "ExacTrac 6.0 is a patient positioning and monitoring system.    Model/catalogue numbers:  20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE  49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3  49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR)  49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC  49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR)  49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR)  49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED  49973B ET DATA PREP / REVIEW SYSTEM  49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT  49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY  49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED",
      "product_quantity": "41 systems (13 systems in US; 28 systems to foreign countries)",
      "reason_for_recall": "When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point.  As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.",
      "recall_initiation_date": "20140224",
      "center_classification_date": "20140425",
      "termination_date": "20170424",
      "report_date": "20140507",
      "code_info": "ExacTrac version v.6.0.3.",
      "more_code_info": ""
    }
  ]
}