  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "There are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List 13000-XX, 13100-XX, or 13150-XX) the potential exists for the GemStar Phase 3 pump to fail to power up while connected to the Docking Station, and 2) when either a GemStar Phase 3 (List 13000-XX, 13100-XX, or 13150-XX)or GemStar Phase 4 pump (List 13086-XX, 13087-XX, or 13088-XX) is used in conjunction with both a Docking Station and an External Battery Pack accessory (List 13073-XX) there exists a possibility that the GemStar pump will display error code 11/003 and give an audible alarm indicating excessive input voltage from the external sources. If the GemStar pump detects what is perceived to be more than 3.6 Volts measured on the external voltage input, the pump will stop the infusion and alarm both with an audible sound as well as a visual alarm.",
      "address_2": "",
      "product_quantity": "26,765 docking stations total (6866 in US)",
      "code_info": "*** US DOCKING STATIONS ***     List Numbers: 13075-04-01, 13075-04-03, 13075-04-05, 13075-04-07, 13075-04-09.     *** FOREIGN DOCKING STATIONS ***     List Numbers: 13075-24-01, 13075-24-03, 13075-24-05, 13075-24-07, 13075-92-09.",
      "center_classification_date": "20140501",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Guam, Puerto Rico, and Internationally to Australia, Austria, Bahrain, Belgium, Canada, Chile, Egypt, France,  Germany,  Greece, Hong Kong, Italy, Jordan, Kuwait, Lebanon, Libya, Luxemburg, Netherlands, New Zealand, Oman, Portugal, Saudi Arabia,  Slovenia, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, and United Kingdom.",
      "state": "IL",
      "product_description": "GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and patient bolus cord.  The docking station is also used for hosting the GemStar pump for general infusions.  List Numbers: 13075-XX-XX.",
      "report_date": "20140507",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-1512-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67802",
      "termination_date": "20170123",
      "more_code_info": "",
      "recall_initiation_date": "20140218",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}