  "results": [
    {
      "country": "United States",
      "city": "Laurel",
      "address_1": "8880 Gorman Rd",
      "reason_for_recall": "Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as \u001cTaller Bridge\u001d power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.",
      "address_2": "",
      "product_quantity": "5,993 (5,080 US; 913 OUS)",
      "code_info": "See Attached Spreadsheat",
      "center_classification_date": "20170320",
      "distribution_pattern": "Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,",
      "state": "MD",
      "product_description": "Giraffe OmniBed  Ohmeda Medical Omnibed- The OmniBed is a combination of an infant incubator ( incubator ) and an infant radiant warmer ( warmer ). Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment",
      "report_date": "20170329",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ohmeda Medical",
      "recall_number": "Z-1509-2017",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "76283",
      "termination_date": "20180713",
      "more_code_info": "",
      "recall_initiation_date": "20161118",
      "postal_code": "20723-5800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}