{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87482",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution.",
      "recall_number": "Z-1507-2021",
      "product_description": "Cardiosave Rescue IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85",
      "product_quantity": "10087 total",
      "reason_for_recall": "There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).",
      "recall_initiation_date": "20210322",
      "center_classification_date": "20210428",
      "report_date": "20210505",
      "code_info": "All Systems.  UDI: 10607567112312  10607567108407  10607567113449",
      "more_code_info": ""
    }
  ]
}