{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Laurel",
      "address_1": "8880 Gorman Rd",
      "reason_for_recall": "Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as \u001cTaller Bridge\u001d power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.",
      "address_2": "",
      "product_quantity": "5,993 (5,080 US; 913 OUS)",
      "code_info": "See Attached Spreadsheat",
      "center_classification_date": "20170320",
      "distribution_pattern": "Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,",
      "state": "MD",
      "product_description": "Giraffe Bedded Warmer, Panda Warmer  Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.",
      "report_date": "20170329",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ohmeda Medical",
      "recall_number": "Z-1506-2017",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "76283",
      "termination_date": "20180713",
      "more_code_info": "",
      "recall_initiation_date": "20161118",
      "postal_code": "20723-5800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}