      "status": "Terminated",
      "city": "Racine",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73685",
      "recalling_firm": "Ad-Tech Medical Instrument Corporation",
      "address_1": "1901 William St",
      "address_2": "N/A",
      "postal_code": "53404-1876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the states of GA and  MI, and the countries of FRANCE, UNITED KINGDOM, COLOMBIA, and  BRAZIL.",
      "recall_number": "Z-1503-2016",
      "product_description": "Ad-Tech Electrode Connection System/Cables.  Lightweight TECH-ATTACH Cable Models:    1.  L-DCL-4DINX (4 contacts:  1 terminal block with 4 connectors each);    2.  L-DCL-6DINX (6 contacts:  1 terminal block with 6 connectors each);     3.  L-DCL-8DINX (8 contacts:  1 terminal block with 8 connectors each);     4.  L-DCL-16BDINX (16 contacts:  1 terminal block with 16 connectors each);     5.  L-DCL-32BDINX (32 contacts:  2 terminal blocks with 16 connectors each);    6.  L-DCL-64BDINX (64 contacts:  4 terminal blocks with 16 connectors each).    FO TECH-ATTACH Connection Cable model FO-LDC-4DINX (4 contacts:  1 terminal block with 4 connectors each).  Tech-Attach:  these cables require the use of an additional component - the connector block.  The tail of the electrode is inserted into the block and the block in turn mates with the cable.  A key-pin is used to ensure proper orientation.  The cable terminated in EEG compatible connector sockets.    Lightweight CABRIO Cable, model L-SRL-64BDINX (64 contacts:  4 terminal blocks with 16 connectors each).  Cabrio cables are similar to Tech-Attach cables except that the block portion is pre-attached.  The tail of an electrode is inserted into this portion which then hinges shut.",
      "product_quantity": "197",
      "reason_for_recall": "Ad-Tech is recalling unopened/unused sterile cables intended to be used only with all compatible Ad-Tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators.  This recall has been initiated due to package failures (i.e. compromised sterile barriers) that were identified during the execution of a shipping simulation investigation.",
      "recall_initiation_date": "20160329",
      "center_classification_date": "20160415",
      "termination_date": "20160824",
      "report_date": "20160427",
      "code_info": "Lot Numbers:  208140507, 208140552, 208140556, 208140559, 208140560, 208140561, 208140562, 208140563, 208140564R, 208140579, 208140581, 208140586, 208140588R, 208140590, 208140604."
    }
  ]
}