  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84841",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide domestic distribution. Worldwide foreign distribution.",
      "recall_number": "Z-1499-2020",
      "product_description": "HeartStart MRx Monitor/Defibrillators with model numbers M3535A  M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5,   M3536M6, M3536M7, M3536M8, and M3536M9.",
      "product_quantity": "97788 units",
      "reason_for_recall": "Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem.  One of these reports involved a death of a patient who could not be resuscitated.",
      "recall_initiation_date": "20200226",
      "center_classification_date": "20200313",
      "termination_date": "20221208",
      "report_date": "20200325",
      "code_info": "All units are affected by this recall",
      "more_code_info": ""
    }
  ]
}