{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65321",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "USA Nationwide Distribution",
      "recall_number": "Z-1495-2013",
      "product_description": "Stryker Instruments 0400-810-000 T4 Zipper Toga  (S/M) Sterile, EO; single use;     Product Usage:  Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.",
      "product_quantity": "7960 units (796 boxes)",
      "reason_for_recall": "Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the   aging of the packaging. Stryker is conducting additional testing and assessing this pr",
      "recall_initiation_date": "20130523",
      "center_classification_date": "20130606",
      "termination_date": "20140414",
      "report_date": "20130612",
      "code_info": "1006129, 1103201, 1009197, 1105391, 1009198, 1107548, 1012302, 1109788,  and 1102127"
    }
  ]
}