  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84756",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution. Foreign distribution worldwide.",
      "recall_number": "Z-1494-2020",
      "product_description": "Cardio Vascular-Allura Centron, system code 722400",
      "product_quantity": "87",
      "reason_for_recall": "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.",
      "recall_initiation_date": "20191230",
      "center_classification_date": "20200313",
      "termination_date": "20221117",
      "report_date": "20200325",
      "code_info": "Not distributed in US.  Serial numbers  132  133  49  66  71  135  31  36  79  69  137  149  147  109  136  108  103  93  53  127  129  129  55  74  2  57  106  56  86  68  157  78  64  63  23  51  100  97  102  113  90  44  30  101  77  83  67  151  22  70  96  116  131  35  82  9  91  43  61  161  28  34  27  19  144  92  168  172  173  122  197  139  6  76  176  16  130  60  80  184  112  81  89  42  110  155  33",
      "more_code_info": ""
    }
  ]
}