  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dover",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82726",
      "recalling_firm": "Zimmer Surgical Inc",
      "address_1": "200 W Ohio Ave",
      "address_2": "N/A",
      "postal_code": "44622-9642",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and countries of AUKLAND  AUSTRALIA  BRAZIL  CANADA   CHILE  Dominican Republic  EL SALVADOR  GUATEMALA  HONG KONG  INDIA    JAPAN               MEXICO            NETHERLANDS                NEW ZEALAND            KOREA          SINGAPORE          TAIWAN                     VENEZUELA",
      "recall_number": "Z-1494-2019",
      "product_description": "Zimmer Biomet Alvarado Knee Holder Base Plate Assembly and Foot Piece, & Alvarado II Base Plate and Foot Piece:  Alvarado Foot Piece:  Item Number: 00-1320-010-00  Alvarado Knee Holder Base Plate Assembly: Item Number: 00-1320-011-00  Alvarado II Foot Piece: Item Number: 00-1320-210-00  Alvarado II Base Plate: Item Number: 00-1320-211-00  Product Usage: This device is designed to hold the patient s knee in the selected position during surgery. The Alvarado Knee Holder includes a Base Plate assembly with two rows of slotted hooks. These rows extend upward from the base plate for a sufficient length to provide adequate selective adjustment. A Foot Piece is mounted on an alignment rod that is pivotally engaged into the hook portion of the slots.",
      "product_quantity": "37,402",
      "reason_for_recall": "Foreign Material in the Alvarado Knee Holder Base Plate Assembly and Foot Piece and Alvarado II Base Plate and Foot Piece due to potentially inadequate cleaning procedures",
      "recall_initiation_date": "20190416",
      "center_classification_date": "20190528",
      "termination_date": "20200519",
      "report_date": "20190605",
      "code_info": "All lot codes"
    }
  ]
}