{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82727",
      "recalling_firm": "Abbott Laboratories Inc. (St Jude Medical)",
      "address_1": "One St. Jude Medical Drive",
      "address_2": "N/A",
      "postal_code": "55117",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA,   SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA,   PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN,   ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA",
      "recall_number": "Z-1493-2019",
      "product_description": "TactiSys Quartz Equipment (with software version 1.7)   Model: PN-004 400   Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)",
      "product_quantity": "317 units",
      "reason_for_recall": "In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data.  This may lead to intermittent contact force data to be displayed during the procedure.",
      "recall_initiation_date": "20190405",
      "center_classification_date": "20190525",
      "termination_date": "20210427",
      "report_date": "20190605",
      "code_info": "UDI:  07640157990194  Serial Numbers:  33019, 33059, 33350, 33354, 33357, 33360, 33361, 33363, 33390, 33392, 33393, 33405, 33417, 33421, 33444, 33449, 33451, 33457, 33458, 33461, 33491, 33492, 33509, 33013, 33015, 33020, 33040, 33075, 33091, 33097, 33160, 33197, 33384, 33385, 33387, 33389, 33414, 33433, 33388, 33435, 33481, 33154, 33157, 33162, 33163, 33278, 33092, 33159, 33437, 33412, 33014, 33027, 33028, 33029, 33038, 33049, 33057, 33082, 33096, 33192, 33222, 33268, 33379, 33383, 33397, 33398, 33399, 33400, 33422, 33423, 33478, 33320, 33352, 33395, 33351, 33069, 33381, 33455, 33064, 33093, 33313, 33337, 33161, 33426, 33083, 33321, 33420, 33051, 33155, 33196, 33256, 33353, 33460, 33217, 33427, 33073, 33150, 33076, 33148, 33434, 33396, 33050, 33333, 33333, 33336, 33336, 33095, 33143, 33143, 33112, 33113, 33266, 33267, 33269, 33270, 33271, 33272, 33273, 33274, 33275, 33276, 33283, 33287, 33288, 33291, 33325, 33356, 33359, 33362, 33404, 33459, 33016, 33098, 33099, 33100, 33101, 33102, 33103, 33104, 33105, 33106, 33107, 33108, 33110, 33111, 33127, 33130, 33131, 33265, 33018, 33024, 33025, 33087, 33152, 33156, 33220, 33230, 33231, 33254, 33279, 33281, 33315, 33416, 33428, 33438, 33439, 33442, 33443, 33463, 33330, 33386, 33228, 33033, 33048, 33149, 33259, 33282, 33464, 33022, 33031, 33035, 33037, 33044, 33045, 33046, 33055, 33066, 33067, 33068, 33078, 33080, 33085, 33086, 33088, 33094, 33129, 33137, 33138, 33140, 33144, 33158, 33195, 33218, 33221, 33229, 33253, 33257, 33262, 33280, 33289, 33293, 33326, 33328, 33355, 33364, 33382, 33403, 33406, 33408, 33409, 33411, 33424, 33431, 33440, 33445, 33446, 33448, 33450, 33454, 33462, 33465, 33466, 33169, 33011, 33012, 33021, 33023, 33034, 33054, 33056, 33074, 33079, 33089, 33090, 33109, 33191, 33193, 33194, 33215, 33224, 33314, 33316, 33327, 33329, 33358, 33375, 33376, 33407, 33415, 33419, 33429, 33447, 33219, 33258, 33292, 33334, 33335, 33032, 33410, 33081, 33170, 33171, 33172, 33173, 33198, 33199, 33201, 33203, 33204, 33205, 33206, 33207, 33208, 33209, 33210, 33211, 33212, 33213, 33214, 33294, 33295, 33296, 33298, 33299, 33300, 33301, 33302, 33303, 33304, 33365, 33366, 33367, 33368, 33369, 33370, 33371, 33372, 33373, 33374, 33549, 33550, 33551, 33168, 33165, 33216, 33042, 33456, 33070, 33432, 33480, 33047, 33072, 33077, 33136, 33226, 33263, 33377, 33378,"
    }
  ]
}