      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).",
      "address_2": "",
      "product_quantity": "67",
      "code_info": "System Serial Number - 76001, 76005, 76007, 76009, 76013, 76014, 76017, 76018, 76019,     System Serial Number - Unique Device Identifier     76058 (01)00884838059450(21)76058  76033 (01)00884838059450(21)76033  76048 (01)00884838059450(21)76048  76059 (01)00884838059450(21)76059  76015 (01)00884838059450(21)76015  76063 (01)00884838059450(21)76063  76066 (01)00884838059450(21)76066  76049 (01)00884838059450(21)76049  76040 (01)00884838059450(21)76040  76032 (01)00884838059450(21)76032  76061 (01)00884838059450(21)76061  76031 (01)00884838059450(21)76031  76006 (01)00884838059450(21)76006  76041 (01)00884838059450(21)76041  76035 (01)00884838059450(21)76035  76038 (01)00884838059450(21)76038  76065 (01)00884838059450(21)76065  76064 (01)00884838059450(21)76064  76053 (01)00884838059450(21)76053  76050 (01)00884838059450(21)76050  76060 (01)00884838059450(21)76060  76036 (01)00884838059450(21)76036  76056 (01)00884838059450(21)76056  76055 (01)00884838059450(21)76055  76037 (01)00884838059450(21)76037  76051 (01)00884838059450(21)76051  76044 (01)00884838059450(21)76044  76039 (01)00884838059450(21)76039  76012 (01)00884838059450(21)76012  76026 (01)00884838059450(21)76026  760016 (01)00884838059450(21)760016  760001 (01)00884838059450(21)760001  760005 (01)00884838059450(21)760005  760004 (01)00884838059450(21)760004  760003 (01)00884838059450(21)760003  760007 (01)00884838059450(21)760007  76072 (01)00884838059450(21)76072  76062 (01)00884838059450(21)76062  76043 (01)00884838059450(21)76043  760013 (01)00884838059450(21)760013  760014 (01)00884838059450(21)760014  76029 (01)00884838059450(21)76029  760010 (01)00884838059450(21)760010  76027 (01)00884838059450(21)76027  760011 (01)00884838059450(21)760011  760009 (01)00884838059450(21)760009  760006 (01)00884838059450(21)760006  76045 (01)00884838059450(21)76045  76054 (01)00884838059450(21)76054  76069 (01)00884838059450(21)76069  76034 (01)00884838059450(21)76034  76030 (01)00884838059450(21)76030  76047 (01)00884838059450(21)76047  76057 (01)00884838059450(21)76057  76028 (01)00884838059450(21)76028  76068 (01)00884838059450(21)76068  760012 (01)00884838059450(21)760012  760002 (01)00884838059450(21)760002",
      "center_classification_date": "20170320",
      "distribution_pattern": "Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.",
      "state": "OH",
      "product_description": "Brilliance BigBore Radiology CT - 728244  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.",
      "report_date": "20170329",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-1493-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76683",
      "termination_date": "20190226",
      "more_code_info": "",
      "recall_initiation_date": "20170223",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}