  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67746",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV,and WY, and the countries of   Austria, Canada, Sweden, England, UAE, Netherlands, Spain, and Hong Kong.",
      "recall_number": "Z-1493-2014",
      "product_description": "Stryker PenAdapt\"; Rx Only, Sterile.    This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing.",
      "product_quantity": "35,310 unexpired units",
      "reason_for_recall": "During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier.  Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the",
      "recall_initiation_date": "20140326",
      "center_classification_date": "20140424",
      "termination_date": "20141211",
      "report_date": "20140430",
      "code_info": "Part Number 0702-045-027",
      "more_code_info": ""
    }
  ]
}