  "results": [
    {
      "status": "Terminated",
      "city": "Louvain La Neuve",
      "state": "N/A",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67678",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "Chemin du Cyclotron, 3",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution including the states of  OK, IL, NJ, and WA.",
      "recall_number": "Z-1492-2014",
      "product_description": "Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.    Proton therapy",
      "product_quantity": "4",
      "reason_for_recall": "Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).",
      "recall_initiation_date": "20131204",
      "center_classification_date": "20140424",
      "termination_date": "20140612",
      "report_date": "20140430",
      "code_info": "Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.",
      "more_code_info": ""
    }
  ]
}