  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87649",
      "recalling_firm": "Ortho-Clinical Diagnostics, INc.",
      "address_1": "513 Technology Blvd",
      "address_2": "N/A",
      "postal_code": "14626-3601",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.",
      "recall_number": "Z-1489-2021",
      "product_description": "VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.",
      "product_quantity": "865 units of 25 slides each",
      "reason_for_recall": "Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides.  If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride.  If the customer doesn\u0019t have any White Reference Slides in reserve, they can\u0019t run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered.  This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.",
      "recall_initiation_date": "20210326",
      "center_classification_date": "20210423",
      "termination_date": "20230524",
      "report_date": "20210505",
      "code_info": "Lots: 9252-0058-3361, 9252-0058-3362",
      "more_code_info": ""
    }
  ]
}