      "status": "Terminated",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67959",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "47215 Lakeview Blvd",
      "address_2": "N/A",
      "postal_code": "94538-6530",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution in the states of NY, NC, DC, FL, PA",
      "recall_number": "Z-1487-2014",
      "product_description": "iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software.  Used for ultrasound examinations of intravascular pathology.",
      "product_quantity": "9 units",
      "reason_for_recall": "Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.",
      "recall_initiation_date": "20140403",
      "center_classification_date": "20140422",
      "termination_date": "20150318",
      "report_date": "20140430",
      "code_info": "Model Numbers:  H749ILAB120CART0  H749ILAB120INS0  H749ILAB120N2710  Serial numbers:  4219, 4900, 5701, 5827  7516, 7517, 7607  8407, 8434.  the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.",
      "more_code_info": ""
    }
  ]
}