  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84756",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution. Foreign distribution worldwide.",
      "recall_number": "Z-1486-2020",
      "product_description": "Allura Exper FD Series. System codes  722003  722010  722022  722005  722011  722019  722001  722002  722006  722012  722008  722013  722020  722025  722015  722023  722038  889006  889016  889009  722005        The device is an X-Ray Generator.",
      "product_quantity": "4938",
      "reason_for_recall": "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.",
      "recall_initiation_date": "20191230",
      "center_classification_date": "20200313",
      "termination_date": "20221117",
      "report_date": "20200325",
      "code_info": "Serial Codes   491  1117  327  316  535  2522  505  1681  133  2268  1001  972  2577  S01H006016/000001  CV149  1651  231  601  1014  2324  1673  CV.563  1712  1043  1806  2229  247  585  217  313  2158  1946  476  1545  517  107  70  492  1891  1715  157  265  935  145  83  71  1356  50  1250  1659  34  2462  1613  2006  480  2562  1550  1440  1734  1733  2201  1922  741  499  2283  1268  1359  833  116  1599  153  787  410  1848  1445  2122  865  360  268  1238  135  1655  2131  489  12  290  1908  74  365  1166  1336  2404  635249  1792  1960  169  196  1113  83  2569  284  39  1195  1461  203  494  839  2138  2141  1953  1956  1106  1096  754  308  580  335  1048  1804  175  343  1011  1497  1473  234  1408  914  255  215  105  662  2382  301  2281  637  917  18204  18158  1764  276  1401  1588  155  38  583  595  594  576  108  159  671  677  590  2302  1279  168  2198  452  2151  321  1436  468  1132  161  670  1025  700  142  1188  27  1034  545  2454  861  1209  1467  789  1226  2134  481  893  1214  290  1187  1455  2085  358  2505  2202  411  2301  1108  401  1541  423  319  1927  1928  2570  1705  2571  2524  926  796  274  2307  67  2200  1798  404  1656  98  399  990  126  371  528  1636  1871  1302  2506  1309  206  1119  47U  141  154  195  2586  1761  951  57  160  27  1652  692  1435  34  1590  1597  147  259  1073  92  509  244  106  59  516  328  346  1797  638  2441  2267  23  281  596  1054  1189  527  1442  461  1099  5    307  1674  S4F220301A001  531  1  72246F  299  823  442  1037  1977  566  72  448  148  1606  924  1730  1936    2310    1610  686  1109  319  1108  1728  460  1294  568  500  435  679  82  1499  459  469  119  254  894  42  49  293  467  302  2536  1422  1077  987  986  297  1537  1112  2461  285  29A  58U  453  46  243  1596  306  116  175  1953  821  13  289  968  1944  1426  1347  112  791  1317  161  1649  1895  1900  1723  878  1084  2221  638  342  69  505  286  2140  591  2496  2311  2222  1515  1803  2484  378  35  1731  1017  1020  636  170  2823014  CV.381  142  1700  946  215  380    1811  214  2523  2573  2267  527  214  402  2171  1690  267  1878  1497  782  427  2068  1602  1694  1839  209  418  1053  2498  1045  932  755  1585  1558  1365  113  344182  1806  2445  25  2126  1148  1181  241  920  2065  113  339  1716  8477  684  1941  1265  1512  331  213  215  389  732  751  809  1324  2276  2086  763  1285  2219  992  1734  149  2390  35  1031  1837  204  521  814  137  815  1120  1095  278  297  86  2204  2366  667  1282  1713  1506  1123  932  2338  1098  2456  524  2449  580  1852  2269  647  1307  218  1858  377  758  1176  463  1598  2425  905  201  172  268  270  2572  250  513  1842  1297  2054  993  67  1743  1747  59  655  325  856  1690  286  507  1430  1407  1657  197  1659  2515  67R  285  641  480  100  690  579  1327  5  196  CV605  695  168  1376  1657  300145495-000165  2530  CV.271  733  565  603  1812  1091  1254  979  1064  55  213  889  2105  2452  504  757  735  1280  2138  518  828  103  320  2271  458  1118  931  357  127  740  1249  542  1033  1223  2265  700  82  577  149  490  1752  1075  1542  2203  2565  424  1321  1396  76986  605  606  359  1724  481  444  445  3  66  374  1  159  253  68A  71A  192  187  72A  1210  2092  1209  834  242  2981  1791  597  205  1068  8426  76378  67  213  1140  1889  1610  950  848  904  244  1002  1754  1022  1023  173  1914  2264  1805  140  89  1650  624  1823  3292  2221  1298  484  483  101  290  615  254  255  52  912  1133  843  117  84  469  190  2150  154  266  892  155  605  604  994  308  1126  133  1478  1262  1688  475  1961  518  519  1545  1732  1062  1090  459  649  650  303  767  194  472  2410  1104  1641  234  2574  196  2625  115  1277  18080  1975  316  2585  959  1461  1335  1173  159  1595  489  150  766  1882  464  56  1293  124  129  1583  422  269  1170  451  625  651  496  599  1128  441  882  491  302  1615  322  18268  158  1946  1047  107  1447  99  556  124  784  1231  351  1344  670  279  82  81    77  297  183  1269  1867  646  648  884  33  58  1660  1616  1083  1846  1807  1924  511  942  61  1714  395  122  101  188  151  161  75  1406  1732  352  1761  492  179  345  1105  472  642  643  2017  124  483  550  1387  465  124  169  1270  377  1201  577  578  2282  1236  1233  1234  575  1235  1314  CV.623  1306  2508  530  2561  2308  2869  2256  1456  155  234  562  698  690  855  178  184  476A  212  2269  71  1078  1216  196  450  306  478  768  609  1140  426  1527  2285  2050  242  1042  1041  1797  1793    309  2347  CV.844  387  454    948  2587    601  176  939  2427  1338  2152  310  1268  498  2430  314  1941  113  520  6  2028  2312  1570  592  2212  1019  309  2003  327  551  98  481  104  1868  2313  1743  369  352  767  2008  997616  1987  2054  2283  572  225  179  1214  28101003  267  2172  2256  2167  2089  510  73  1239  1561  10439  1854  170  1960  1508  77  616  44  1084  332  79  78  675  2567  2289  69  18299  308  46  102  780  1038  1058  159  2250  239  639  669  671  318  1704  1366  1442  1990  1449  2568  928  01H041985  81  1026  1941  1337  264  926      427  685  141  1643  72  2112  2111  1622  1621  41  656  1560  1441  2926  134  2288  270  78  525  20343027  354  1035  1209  1818  418  145  830  1799  2263  473  1082  2068  624  CV.718  252  873  395  1586  435  141  1710  2065  79  175  1617  1336  274  1354  1918  619915  1906  818  1904  730  383  560  74  68  644  177  29-A  754  1828  1907  424  90  2351  1405  634919  182  1757  114  1948  1195  518  888  1230  1329  1992  323  379  394  792  791  1263  1525  69  1003  1219  357  2320  102  89  115  132  1064  2326  999  2004  2003  1118  325  15  12  40  85  43  135  100  2525  1832  2133  728  2166    562  216  479  485  165    2308  54  244  298  307  760  488  280  859  1306  1310  673  683  1593  1411  535  315  1608  79  76  637  1950  1947  890  232  641  731  611  2296  170  64  1328  848  2292  520  1972  274  439  1063  17  1587  136  173  2303  1040  308  174  1226  360  613  1271  205  1943  1945  1949  456  392  105  2147  1799  1950  523  718  1647  1695  131  187  1248  109  72  325A  1502  2201  218  228  1727  24    87  1513  304    567  1909  2287  72  1028  2015  2207  1223  76955  2239  2516  1507  1911  425  2309  237  319  1589  1958  2075  209  500  2187  797  1922  389  2161  1729  15  1665  317  1116  252  1380  364  110  61  1993  1844  1443  302  150  49  1281  794  1945  2276  1720    2293  2286  110  39  337  479  1735  1326  95  423  788  561  153  521  2578  262  1118  2143  488  2002  132  1955  1218  154  45  191  2566  903  336  694  691  1405  1102  123  1334  2013  1954  2560  2563  1240  2202  129  1141  66  50  372  797  923  925  1258  533  406  878  778  777  1293  1081  1957  513  65  920  1232  1446  1675  1625  515  2107  2336  2332  606  1683  207  1005  1275  1292  1332  64  267  1305  790  2123  38925  273  976  68  1144  1618  1687  365  1847  1787  453  633  499  19  95  1149  2127  2459  2513  2297  77  934  1910  1726  2261  1916  85  117  753  476  1036  171  1599  485  227  1045  1531  109  1809  447  277  1923  792  1630  851  1827  2443  519  208  943  404  984  1516  453  321  330  401  8095  100089369/2868  334  2184  2355  1716A  197  1869  2215  1283  160  8  902  1197  165  2386  574  1611  497  2489  38532  170  61  1707  2257  2099  362  229  565  989  767  111  2064  1559  1869  2925  2924  CV.514  1676  240  1635  593  65  651  1686  1224  858  197  379  788  1345  2380  2272  86  351  296  42  2057  1519  124  1430  1511  214  1255  84  1312  2346  1599  125  81  610  134  56  838  2147  666  1989  900094347  424  27  1183  2022  2014  138  611  1126  1943  1549  1188  921  939  1404  1073  1714  1664  103  1272  564  2063  1662  1711  2319  505  148  1518  970  2350  18  326  693  697221  675  1692  1538  256  2575  298  1217  2936  756  1027  424  585  1894  2148  2145  2146  573  91A  495  826  1244  1185  2518  1731  750  1756  2088  511  197  2227  82  83  1177  2203  93  95  41  1167  2199  98  152  42  40    2135  88  245  2378  2317  1672  1100  909  1210  100  965  633056  168  495    221  1101  500  1846  787  55  2273  2183  445  1399  128  127  513  123  126  136  96  45  515  1328  472  676  31  22  493  75  487  1575  474  497  105  1661  428  275  1005  274  954  953  952    2156  18  1279  427  320  100  186  3KHJ01J  43090  80320  8198  344  455  531  324  570  656  668  973  1115  1854  CV.643  1622  226  1805  22  1078  316  1212    140  1181  2576  1032  278  975  976  347  170  163  986  896  490  386  971  1594  233  239  342  531  640  2081  375  1316  1645  199  1746  470  2305  47  1044  1477  246  62  44  143  2319  43  2564  277  1114  437  337  6001154  192  1551  722008-358  488  1947  226  314  477  4  1236  852  224  1016  60  1097  1855  457  315  1532  122  480  1964  2135  695  CV.740  1790  857  1603  1807  1357  2159  589  1788  474  41  467  269  1471  1092  304  602A  1261  214  191  937  1210  248  241  846  1446  CV603  2412  1942  82  147  963  1009  2148  1533    964  285  15  322  1718  61  1658  422  412  431  694  177  1150  1644  2274  107  42  2  75  414  714  785  1344  84  339  1919  351A  69  1719  2234  1301  462  220  1358  689  1286  2323  33  57  456  430  1542  76  1607  2197  26  456  344  2454  172  2149  1211  249  1004  397  2046  132  974  1069  1287  1308  109  2254  1192  2323  118  1507  26  1432  306  205  2136  444  348  160  543  213  35  25  2  1796  72  446  1902  154  601  215  2424  C@24226  1909  40  405  654  33  1122  146  2286  1132  1361  104  607  101  2078  1198  328  339  781  332  1504  1433  157  229  118  142  962  52  1047  777  1260  567  2108  1983  666  1094  301  489  837  113  1125  333  30  1838  283  1681  608  1337  1052  1081  2282  2313  2337  426  924  1815  922  1333  774  1738  1749  577  265  2047  335  335  1553  168  1492  120  1637  1260  2318  1187  171  142  1655  1855  219  265  959  1817  309  1052  1994  901  1428  1562  627  2005  552  1956  930  64  2168  910  1168  967  167  888  1008  2098  935  2304  1140  931  1421  1802  2371  1213  1229  1015  1016  1017  1018  1369  841  509  1827  1194  2001  15  121  229  232  502  416  994  1674  28  332  374  632  1  1192  137  1035  122  479  2335  524  957  272  273  361  1335  2235  1225  522  475  863  1153  289  2254  1040  2487  496  503  1444  118  2232  409  2193  2258  47  9  1420  105  48  1853  295  721  1188  1189  1208  1190  233  1211  112  418  533  1072  70  1332  1116  1168  2042  821  1675  57  806  995  50  993  907  221  330  119  887  600  326  2231  71  516  400  2350  825  25  1751  759  350  1339  1865  123  1351  1460  195  289  1199  323  1137  1802  508  713  1872  174  1971  173  550  405  664  91  53  948  2107  537  914  1148  502  1702  248  1684  1307  596  45  1593  2389  2352  187  1304  1156  180  1197  159  2314  1464  347  2335  137  795  628  1001  92  602  1346  619  139  134  1743  809  930  109  179  575  576  170  302  2406  66  231  230  232  991  589  245  509  1093  1349  1124  176  1124  1376  444  1291  535  370  2169  553  941  2244  263  712  47  1726  402  1712  482  1363  1457  1034  25  2020  288  553  393  462  1980  203  1580  447  2236  292  395  400  487  1485  291  2360  1256  342  2416  876  1592  582  1222  2346  2188  2325  2074  2072  805  1076  4.5161E+15  392  206  2457  992  60  477  381  2333  1480  1986  1750  750  1228  328  2076  2071  1331  2032  630  1196  98  447  1495  2485  1614  2139  1591  1357  2321  261  1966  1825  1839  1054  1385  1582  215  2490  2204  598  513  1861  218  2375  169  1413  1122  608  541  642  2178  978  2356  166  746  652  2411  665  1678  2029  2413  114  755  1435  374  2451  410  1081  770  1298  2392  1028  2214  1349  152  2450  2056  911  1366  1708  117  2082  1059  1984  830  275  2479  1626  2403  599  1572  2173  2174  1978  1808  548  2455  1244  829  2363  393  228  69  891  1899  2102  2300  2181  181  1668  2297  663  1434  1076  2398  839  1348  817  91  1814  2298  572  2182  1201  1100  1882  569  2399  2391  2392  688  2255  2016  1651  544  1481  338  1390  2325  419  1699  2197  407  1436  2336  2296  728  1296  785  391  355  1698  2456  803  705  1556  436  2386  1214  2349  2418  1736  1397  1967  2096  210  1691  2364  2366  2291  2299  622  1169  1223  58  128  1878  2122  1937  1669  2423  1623  648  382  1471  737  2388  1879  581  2121  47  112  2124  1194  1818  1150  941  856  867  540  993  1816  1288  1489  1178  1874  2349  2175  479  211  802  803  1303  1147  2176  866  2208  1175  1237  1381  555  2274  2492  398  1267  2459  2252  1765  657  658  659  854  1071  271  1191  1965  2427  1360  798  441  1708  1709  2026  2412  706  1089  707  708  1145  1144  369  2272  2429  1342  1828  1535  2393  1463  647  2196  1479  534  2421  116  165  85  1203  2488  870  2344  1536  1654  2249  699  2394  2153  748  2094  2410  772  2120  660  1739  1021  1382  2326  539  1835  2345  2034  1999  1285  2405  1450  1309  1483  2109  1836  2343  680  2481  980  2058  2118  2358  51  1142  916  360  1263  74  1350  2396  1702  2340  730  2031  779  1976  732  1079  296  987  2127  2433  1064  1917  1287  1795  1137  704  2401  1412  944  454  613  614  635  742  238  227  506  2378  1342  509  1671  667  1929  2287  280  279  125  432  966  928  2380  2252  2376  1631  1484  2379  752  2387  2104  1875  2443  2348  1332  216  1113  2438  2442  2361  1830  1411  2444  559  1127  2330  659  2314  528  1740  1711  813  1624  593  2128  2192  1898  2460  1297  29  1701  2341  1220  2179  2067  1206  2467  1061  443  405  2404  289  1254  587  1253  1143  260  1032  1530  903  1252  1364  2327  748  1233  219  1360  538  2290  513  2030  2482  1234  1284  164  1151  1152  2117  2017  106  524  2437  199  192  1833  1165  1232  1258  2466  1561  1860  478  1705  1721  215  871  506  793  794  1423  753  2244  2372  2373  266  1462  507  400  287  1819  727  743  799  1215  2475  1348  1347  1977  490  763  1703  2351  580  1256  1998  693  1653  388  311  554  902  502  1289  199  2348  1130  264  1548  2384  2434  1887  1567  1530  2470  59  2270  917  247  350  1241  2061  2371  1958  2342  2278  661  759  2250  810  771  998  2417  2477  1131  2155  722  1174  2205  117  546  620  1375  982  1268  965  2101  662  424  556  2487  2033  2288  1885  2123  2292  2399  1163  1164  56  2331  1997  2431  129  1537  324  2389  2028  2112  1361  979  2338  573  695  1128  696  2436  414  2007  2291  1135  955  214  1138  936  2390  2453  1088  1927  1086  425  565  588  2087  384  2385  269  679  971  2103  2449  2237  895  272  376  2367  1384  1377  391  2301  1242  1695  2334  374  738  1255  619  2502  1366  1321  469  869  105  1415  735  736  734  737  1202  335  764  185  2434  2080  1693  2422  1223  1353  222  2307  731  1019  1023  541  57  1881  1098  1099  390  2008  2200  415  2095  1266  2094  1056  1364  733  1093  2369  1327  773  868  2393  1134  2024  2414  2415  1862  2305  2439  806  629  1680  1741  1893  184  61  1232  1737  901  1356  332  1735  898  2160  319  2278  2177  1367  2432  1355  2195  416  1801  1388  1352  726  820  188  2140  2489  1391  1972  223  445  2245  2093  1344  1325  127  1157  833  1834  949  1926  2198  2293  1717  631  415  1074  2073  1996  2382  1043  603  361  1003  1261  1855  175  176  476  164  557  222  840  437  1752  1082  313  340  426  303  1242  392  270  267  268  776  376  1633  217  348  47  325  672  358  1089  422  2  44  885  1579  1379  872  549  476  2428  293  294  692  144  530  272  1312  831  286  1821  1820  133  2077  1822  1973  81  1091  531  224  104  390  1204  135  442  505  161  1677  940  1389  34  29  63  578  1453  1008  884  368  1078  1010  2453  197  366  1668  1095  322  49  273  118  371  146  70  346  2374  1837  33  315  484  1959  1688  2539  73  74  11  73  126  772  985  5    1696  2247  119  59  2045  1746  1747  2043  1826  621  970  622  56  193  412  678  36  62  483  1820  1161  335  1565  780  314  25  2240  155  419  1589  1590  1591  396  5  140  1917  31  35  173  786  225  518  1665  154  1466  31  1674  984  616  1292  160  53  908  534  2218  1290  34  3067  310  1091  762  2414  615  360  380  CV.673  1918  1225  1224  685  686  1931  1086  2547  46  2368  390  8  312  1912  1772  2538  1969  17  78  175  1906  1585  4  1345  2550  1151  1770  452  117  16  776  2206  14  117  334  338  1002  799  711  177  946  1087  1774  1697  1540  807  598  1885  2304  58  1166  1245  2551  1927  676  984  990  1930  CV.676  72  1930  1125  110  232  1916  CV.494  376  1716  520  146  2584  23  82  6  957  1926  1897  1245  556  1905  195  331  2144  778  1860  771  299  1036  773  1939  1004  1013  1246  282  1494  CV.502  761  1268  544  327  1925  72  478  174  2196  158  880  715  1158  1928  1951  897  1221  2137  2419  716  718  1010  1241  1  222  523  145  21  1936  1436  2439  2548  63  599  355  1104  74  1663  2921  1824  231  323  1753  129  2554  1153  224  1159  363  847  978  609  663  664  2000  345  1269  911  1400  617  CV.698  295  691  1889  1169  288  417  1880  1908  1798  543  560  403  80  10  109  1759  26  1399  653  1823  164  417  551  1510  407  2526  2426  1083  490  835  1451  586  107  1134  1135  2552  2465  2541  25  151  2529  499  557  148  1937  1429  1495  863  253  385  1748  2555  635  184  90  9  1935  1180  1632  86  889  828  918  398  289  49  48  1845  41  86  13  266  1131  CV.705  21  528  1425  1988  CV.265  418  1923  561  542  276  819  1929  80  203  207  111  202  2549  351  105  1653  5  740  560  355  8  165  53  2238  525  238  281  777  167  179  157  625  749  247  1720  CV.748  CV.719  2499  936  1546  CV.696  1253  618  26  2514  48  68  631  163  1155  393  372  592  1303  2285  1346  37  367  453  224  76  447  372  210  203  214  369  613  1870  131  620  963  496  287  478  34  35  299  915  471  597  80  510  1723  1773  162  102  739  128  1315  401  795  544  1178  1177  811  600  1838  649  172  692  1050  498  540  655  159  894  961  962  382  798  769  28  942  246  1157  610  1755  1756  839  623  409  497  496  495  3039  1238  136  1150  944  955  544  52  339  650  2316  300  2220  1138  751  1817  1111  514  232  249  1196  353  903  1141  3298  1022  164  877  255  361  495  1272  384  598  1058  521  801  498  797  549  1985  1145  530  1730  1030  574  1105  1276  1049  687  336  349  1278  97  415  1915  1818  154  1216  897  1712  1274  1273  516  100  416  1169  1037  1784  1191  954  866  585  2079  969  311  99  50  652  1294  1106  608  1039  1654  526  674  88  1829  66  410  758  1358  1090  1710  1012  703  1265  860  86  1202  798  2186  1154  1987  1704  494  275  2294  1392  967  927  1286  861  602  475  775  2934  356  876  1722  1250  1574  293  37  139  1897  612  1007  1305  1306  1203  271  83  1834  895  800  359  375  383  872  98  906  680  6  1033  340  570  733  428  960  1793  427  261  1021  355  3  983  943  743  951  832  474  1266  1237  333  90  720  1172  457  1322  945  406  2206  745  1891  1121  543  837  1120  2189  1979  547  132  808  404  928  1053  428  1719  579  503  493  465  2343  1024  988  2411  824  824  1582  807  1048  529  211  217  351  617  528  559  1334  1682  88  527  138  2  985  58  CV619  11  167  168  107  14  138  566  569  900  588  589  590  591  593  594  58  891  949  1146  1267  1534  6000661  199  354  328  75  245  122  2310  2309  1823  998  1527  1170  341  225  983  73  2295  250  1101  518  568  70  1243  385  1350  58  163  63  168  1857  1259  138  1764  280  179  885  29  2437  751  1620  264  696  262  784  227  508  1121  2275  1225  659  3  220  421  904  700  701  1136  2927  538  2011  106  1692  810  850  76  802  318  2918  103  104  546  312  1180  522  109  595  913  956  1452  194  1059  198  1097  842  1813  2353  77  744  2429  182  294  621  354  347  574  337  181  98  498  519  103  13  569  2266  2559  523  482  339  429  2556  292  136  92  208  408  246  1850  2241  202  516  207  356  355  168  42  406  126  475  576  1609  1955  2922  1634  314  542  58  178  2499  555  1634  388  94  529  972  2059  486  547  127  418  961  131  452  295  493  180  2594  551  141  404  1251  207  1300  2375  533  514  933  1076  568  550  157  1198  52  211  411  290  291  1553  1520  1424  858  273  2280  275  1701  321  622  1015  131  261  90  65  230  1840  1077  38  39  176  1939  1541  2493  710  1418  517  518  1961  32  554  317  524  249  2597  395  520  12  212  2600  754  204  37  521  2185  510  348  503  26  259  466  271  450  555  134  541  9  216  338  263  160  535  49  2505  2593  94  1968  522  2458  956  1154  2210  193  334  55  131  102  72  1250  1867  45  1547  586  874  120  681  190  394  1966  514  545  1949  257  1846  2141  502  457  465  760  104  2497  816  1964  113  1176  1742  56  64  343  192  53  595  1218  1230  790  219  563  2557  1055  70  150  564  546  2595  1676  2379  494  2367  563  127  55  98  292  536  409  99  28  57  587  105  2207  2209  294  887  1207  292  345  1901  126  458  1639  244  1368  225  226  125  60  2216  501  552  540  572  531  76  396  149  133  268  257  265  859  2639  198  1619  282  250  1739  21  1454  6  549  1899  264  20  7  2663  1744  492  1338  217  69  20  403  571  1962  1130  425  573  109  45  189  416  271  1280  940  85  434  139  362  258  2066  361  79  934  845  128  1313  864  340  1812  378  241  249  412  1963  1940  486  336  1131  344  206  353  56  52  2090  183  1124  645  6001385  2598  201  465  436  522  329  855  221  488  537  532  526  135  553  333  475  2100  95  131  1149  503  68  84  201  1691  178  114  575  853  123  1156  557  67  253  2498  2005  119  1962  178  145  442  283  296  367  158  501  222  287  137  551  152  171  56  112  276  377  378  376  11  138  318  2110  637  182  201  53  341  473  75  38  1112  1042  135  78  144  2625  1417  261  1543  289  197  143  2558  657  15  567  474  219  62  366  2036  527  558  468  236  86  1182  324  1082  438  276  493  504  1227  311  459  606  70  116  1931  2259  161  320  247  303  1060  514  88  1651  2217  449  491  932  439  321  435  545  300  176  65  499  1689  997  480  309  181  554  925  1952  4  139  370  257  1554  1200  23  226  950  24  1051  1647  1065  440  504  464  439  1021  525  1341  2300  1932  952  548  1123  235  104  294  181  526  292  167  1220  312  753  1935  114  815  1763  130  1535  455  115  769  562  180  184  183  185  186  187  188  189  190  529  413  437  303  302  304  301  305  306  307  408  6000296  789  1067  1847  1129  2089  873  874  975  99  9  233  31  173  89  523  156  994  1724  484  71  727  325  1046  1745  508  481  222  539  626  77  236  188  108  243  1700  2520  1991  853  2084  1313  1180  160  2475  186  2362  342  6131  2359  2129  1504  2342  2581  1200  1355  341  470  1509  2361  2360  786  1925  1350  83491  373  2097  744  100  223  1648  212  440  1804  555  598  1171  1551  702  1794  1119  2091  1111  471  144  223  1051  989  2163  28  1615  149  35  189  854  480  164  340  953  927  210  1676  312  338  137  1080  458  2460  347  618  682  321  399  238  1439  95  1844  CV.389  966  34  1911  1748  464  463  NL94  660  1697  2431  54  385  741  742  1029  1088  26  805  1886  1136  353  153  51  2114  403  500  304  881  237  91  95  844  1283  192  973  81  448  497  1448  1206  267  2154  1123  228  1162  1974  11  2589  85  890  142  2230  30  425  2253  1330  421  87  94  7567201  1050  111  621  818  394  1888  2334  561  57  1000  1474  1907  1171  130  948  277  134  269  919  1581  764  177  381  1374  382  162  1747  1744  1760  68  129  304  101  558  155  822  1031  1371  73  1864  880  405  862  58  804  1044  1351  132  1517  945  1318  836  1085  380  889  1087  946  179  2078  1565  225  487  226  476  1264  2290  1518  1186  645  1557  1566  1125  1372  374  1449  852  256  538  1637  1574  1640  278  1575  1576  1573  1248  1493  60  458  1577  119  118  1667  468  1578  367  455  646  191  301  302  186  352  123  62  410  32  108  101  41  55  1579  108  300  320  299  96  71  1403  1402  89  263  258  138  99  2069  1914  311  310  165  396  20  832  239  654  371  329  2341  470  87  1173  233  86  474  66  274  2311  117  722008530  46  373  169  129  254  506  139  243  141  124  918  968  27  202  2211  30  626  487  1684  6  2580  1117  378  165  304  305  110  306  307  111  308  1150  1951  488  363  3354  330  367  473  515  909  1569  220  16  2531  433  343  373  875  2280  112  895619  580  2528  S4F002709A  413  1152  2532  563  14  429  2527  2243  644  900  5  327    1159  50  765  312  313  158  28  133  51  2517  1129  PP0200000118  2010  1490  1045  1046  2933  121  124  1075  1074  413  807  298  44  840  564  362  1193  754  762  1544  30  115  1670  20  247  246  55  793  1469  357  921  1066  125  636  1669  511  135  194  1092  89  97  5  86  434  3  227  834  2347  882  1391  243  1920  191  482  96  184  3461  296  237  427  1051  567  100  533  453  165  220  242  166  299  507  364  543  273  190  189  532  475  260  280  74  75  422  1304  857  128  251  228  541  161  194  542  697  291  185  250  91  314  199  223  343  634  22  1057  1853  980  269  1185  1295  230  1055  1779  783  782  213  1096  675  103  128  108  212  849  1327  12  27  3  125  661  1008  481  1850  2242  225  812  813  370  1778  193  368  913  336  1733  349  2055  115  354  2194  964  996  2037  463  781  235  236  1706  152  307  111  549  850  163  468  2225  388  1090  387  603  113  386  526  1968  1129  236  525  1841  112  132  1959  114  66  570  116  604  283  1302  283  868  460  454  251  278  774  157  1468  10  1006  2289  156  1060  558  528  2391  516  517  812  2053  1398  665  1627  50  601  1555  1628  2373  156  1868  2164  2273  467  403  1666  1896  1030  1031  43  597  368  1921  492  908  845  846  507  923  277  313  687  1320  1404  177  2114  140  141  303  1212  1  607  847  817  2854  293  560  1969  2130  106  183  375  381  54  60  838  2228  2226  2219  2220  26  27  29  98  775  78  1528  1503  499  166  501  770  1056  46  719  1982  1781  2248  240  1706  185  67  1810  1127  130  129  31  969  411  1715  395  634  295  39  242  241  157  156  1008  1341  804  658  1068  443  540  1432  491  259  1147  ERROR1008  2279  689",
      "more_code_info": ""
    }
  ]
}