{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dover",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65258",
      "recalling_firm": "Zimmer Surgical Inc",
      "address_1": "200 W Ohio Ave",
      "address_2": "N/A",
      "postal_code": "44622-9642",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.",
      "recall_number": "Z-1484-2013",
      "product_description": "Zimmer Air Dermatome II, REF 00-8851-001-00  handpiece  Rx, non-sterile    grafting instruments intended to provide variable graft thickness and width capabilities",
      "product_quantity": "126 units",
      "reason_for_recall": "This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken.  This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.",
      "recall_initiation_date": "20130507",
      "center_classification_date": "20130605",
      "termination_date": "20141215",
      "report_date": "20130612",
      "code_info": "Serial Number 500191 , 500192,500193, 500194,500196,500197,500198,500199, 500200, 500201,500202,500203, 500204, 500205,500206, 500207,500208,500209,500210, 500211,500212, 500213,500215,500216,500217, 500219,500220, 500221,500222, 500223,  500224, 500225,500226,500227,500228,500229,500230, 500231,500232, 500233,500234,500235,500236, 500237, 500238,500239,500240,500241 ,  500242, 500243,500244,500245,500246,500247,500248,500249, 500250, 500251,500252,500253,500254, 500255,500256,500257,500258,500259,  500260,500261,500262,500263,500264, 500265,500266,500267,500268,500269, 500270,500291,  500292,500293,500295,500296,500297,500298,  500299,500300,500301 , 500302,500303, 500304,  500306, 500307,500308,500309,500310, 500331 ,  500332, 500334, 500335, 500336, 500337, 500338,  500339, 500340, 500341,500342, 500343, 500344,  500345, 500346, 500347, 500348,500349, 500350,  500351 , 500352, 500353, 500354,500355, 500357,  500358, 500359, 500368, 500369,500370,500371"
    }
  ]
}