  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Milwaukee",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87686",
      "recalling_firm": "Welch Allyn Inc Mortara",
      "address_1": "7865 N 86th St",
      "address_2": "N/A",
      "postal_code": "53224-3431",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, CO, FL, GA, IN, LA, NC, NJ, NV, NY, OH, OK, TX, UT, and VA.",
      "recall_number": "Z-1483-2021",
      "product_description": "ELI 380 Electrocardiograph, with AM12M Acquisition Module accessory.  Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.",
      "product_quantity": "27 units",
      "reason_for_recall": "Devices do not meet IEC 60601-2-27 requirements as labeled.",
      "recall_initiation_date": "20210331",
      "center_classification_date": "20210422",
      "termination_date": "20240816",
      "report_date": "20210428",
      "code_info": "Part Number: ELI380-ACX32 (Device Identifier: 732094265415); S/N: 118490001284, 119150000588, 120090001355, 120090001354    Part Number: ELI380-DAX3X (Device Identifier: 812345026757); S/N: 117120557018, 117120557019, 117120557020, 117120557021, 117230571942, 117230571943, 117270577469    Part Number: ELI380-DBX32 (Device Identifier: 732094265262); S/N: 118180634420    Part Number: ELI380-DCX32 (Device Identifier: 732094265187); S/N: 118340001074, 118340001074, 118340001075, 118500001332, 118460001178, 119150000120, 119210000774, 119290000580, 119520000194, 120080001509, 120080001510, 120100000310, 120080001511, 120110002478, 120170000570",
      "more_code_info": ""
    }
  ]
}