  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79773",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.",
      "recall_number": "Z-1483-2018",
      "product_description": "Fabius Tiro, anesthesia machine, catalog no. 8606000    Product Usage:  The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).",
      "product_quantity": "26",
      "reason_for_recall": "The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.",
      "recall_initiation_date": "20180301",
      "center_classification_date": "20180507",
      "termination_date": "20190416",
      "report_date": "20180516",
      "code_info": "Serial no: ASKJ-0195  ASKK-0079  ASKF-0226  ASKF-0227  ASKF-0228  ASKF-0229  ASKF-0230  ASKH-0086  ASKH-0071  ASKJ-0177  ASKK-0020  ASKK-0021  ASKH-0053  ASKH-0054  ASKH-0055  ASKH-0056  ASKH-0057  ASKH-0067  ASKH-0068  ASKH-0063  ASKH-0065  ASKH-0064  ASKH-0130  ASKH-0131  ASKJ-0173  ASKH-0031"
    }
  ]
}