  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79773",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.",
      "recall_number": "Z-1482-2018",
      "product_description": "Fabius GS Premium, anesthesia machine, catalog no. 8607000    Product Usage:  The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).",
      "product_quantity": "27",
      "reason_for_recall": "The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.",
      "recall_initiation_date": "20180301",
      "center_classification_date": "20180507",
      "termination_date": "20190416",
      "report_date": "20180516",
      "code_info": "Serial no: ASKF-0224  ASKF-0232  ASKH-0004  ASKH-0007  ASKH-0038  ASKH-0088  ASKH-0132  ASKH-0133  ASKJ-0007  ASKJ-0178  ASKJ-0179  ASKJ-0184  ASKJ-0205  ASKJ-0220  ASKJ-0224  ASKK-0002  ASKK-0006  ASKK-0010  ASKK-0011  ASKK-0014  ASKK-0019  ASKK-0022  ASKK-0023  ASKK-0032  ASKK-0036  ASKK-0038  ASKK-0044"
    }
  ]
}