{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92229",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution",
      "recall_number": "Z-1481-2023",
      "product_description": "LUMINOS Lotus Max (VF11)",
      "product_quantity": "704 units in the United States (1945 units worldwide)",
      "reason_for_recall": "Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.",
      "recall_initiation_date": "20221123",
      "center_classification_date": "20230523",
      "report_date": "20230531",
      "code_info": "Model: 11574100"
    }
  ]
}