{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79773",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.",
      "recall_number": "Z-1481-2018",
      "product_description": "Fabius MRI, anesthesia machine, catalog no. 8607300    Product Usage:  The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.",
      "product_quantity": "8",
      "reason_for_recall": "The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.",
      "recall_initiation_date": "20180301",
      "center_classification_date": "20180507",
      "termination_date": "20190416",
      "report_date": "20180516",
      "code_info": "Serial no: ASKM-0020  ASKM-0011  ASKL-0005  ASKL-0007  ASKL-0010  ASKM-0031  ASKM-0032  ASKM-0023"
    }
  ]
}