  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87647",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.",
      "recall_number": "Z-1477-2021",
      "product_description": "Stopcock rotator (Component):  a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR  b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL  c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R.  Stopcocks and manifolds used to control the direction of IV fluid flow.",
      "product_quantity": "3825 units",
      "reason_for_recall": "Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.",
      "recall_initiation_date": "20210331",
      "center_classification_date": "20210421",
      "termination_date": "20230320",
      "report_date": "20210428",
      "code_info": "a. Model Number MX1431MR, Lot Numbers:  3995544  b. Model Number MX1431MRL, Lot Numbers:  3995545, 4011838, 4027758, 4033501, 4044959  c. Model Number MX4331R, Lot Numbers:  3971467, 3987918",
      "more_code_info": ""
    }
  ]
}