  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87647",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.",
      "recall_number": "Z-1476-2021",
      "product_description": "Manifold rotator (Component):  a. 250 2 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1232MR  b. 250 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1233MR  c. 550 2 GANG MANIFOLD ROTATOR, OPEN HANDLE, Model Number MX1322MR  d. 550 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1333MR. Stopcocks and manifolds are used to control the direction of IV fluid flow. component of extravascular  blood-pressure Transducer.",
      "product_quantity": "1380 units",
      "reason_for_recall": "Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.",
      "recall_initiation_date": "20210331",
      "center_classification_date": "20210421",
      "termination_date": "20230320",
      "report_date": "20210428",
      "code_info": "a. Model Number MX1232MR, Lot Numbers:  4027757  b. Model Number MX1233MR, Lot Numbers:  4076456  c. Model Number MX1322MR, Lot Numbers:  3992363, 4011839  d. Model Number MX1333MR, Lot Numbers:  4016439",
      "more_code_info": ""
    }
  ]
}