  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rillieux La Pape",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84928",
      "recalling_firm": "Medicrea International",
      "address_1": "5389 Route De Strasbourg",
      "address_2": "Vancia",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.",
      "recall_number": "Z-1473-2020",
      "product_description": "Iliac Screws, multiple sizes,   Article Nos. B36020500  B36020570  B36020580  B36020590  B36027500  B36027570  B36027580  B36027590  B36028500  B36028570  B36028580  B36028590  B36029500  B36029570  B36029580  B36029590    Screws manufactured in titanium alloy Ti-6Al-4V LI conforming to ISO5832-3 and ASTM.   F136 specifications. These screws are part of an internal fixation system composed of hooks, rods, plates, cross links, connection and locking devices.",
      "product_quantity": "3529",
      "reason_for_recall": "An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.",
      "recall_initiation_date": "20200128",
      "center_classification_date": "20200311",
      "termination_date": "20230317",
      "report_date": "20200318",
      "code_info": "Lots 19D0683  19D0684  19D0685  19D0686  19D0369  19H0470  19D0370  19H0473  19D0371  19G1071  19L0501  19D0372  19G1073  19D0570  19G1075  19J0445  19D0571  19G1076  19J0867  19D0572  19G0521  19J0868  19L0526  19L0502  19L0525  19D0573  19G0522  19J0869  19D0574  19G1077  19L0503  19D0575  19H0474  19D0576  19G0523  19G1078  19D0577  19G1079  19H0475",
      "more_code_info": ""
    }
  ]
}