      "status": "Terminated",
      "city": "Rillieux La Pape",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84928",
      "recalling_firm": "Medicrea International",
      "address_1": "5389 Route De Strasbourg",
      "address_2": "Vancia",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.",
      "recall_number": "Z-1472-2020",
      "product_description": "Pedicle Screws, multiple sizes,    Article Nos. B36004525  B36004530  B36004535  B36004540  B36004545  B36004550  B36004555  B36005530  B36005535  B36005540  B36005545  B36005550  B36005555  B36005560  B36006530  B36006535  B36006540  B36006545  B36006550  B36006555  B36006560  B36007535  B36007540  B36007545  B36007550  B36007555  B36007560  B36008535  B36008540  B36008545  B36008550  B36008555  B36008560  B36009535  B36009540  B36009545  B36009550  B36009555  B36009560    Screws manufactured in titanium alloy Ti-6Al-4V LI conforming to ISO5832-3 and ASTM.   F136 specifications. These screws are part of an internal fixation system composed of hooks, rods, plates, cross links, connection and locking devices.",
      "product_quantity": "3529",
      "reason_for_recall": "An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.",
      "recall_initiation_date": "20200128",
      "center_classification_date": "20200311",
      "termination_date": "20230317",
      "report_date": "20200318",
      "code_info": "Lots 19C0644  19J1008  19C0645  19J1009  19C0646  19G1063  19J1010  19C0647  19G1065  19J1011  19C0648  19J1012  19C0649  19J1013  19C0650  19J1014  19C0269  19C0269R  19H0458  19C0270  19C0270R  19H0459  19J0440  19L0471  19C0271  19C0271R  19D0219  19H0460  19J0441  19L0205  19L0472  19C0272  19D0220  19C0272R  19H0463  19J0442  19L0206  19C0273  19D0221  19C0273R  19H0464  19J0443  19C0274  19C0274R  19D0222  19C0275  19C0275R  19H0467  19B0847  19B0847R  19L0209  19L0473  19B0848  19D0223  19B0848R  19F0226  19L0214  19B0849  19D0224  19B0849R  19F0227  19L0217  19L0475  19B0851  19D0225  19B0851R  19E0637  19F0267  19G0734  19H0468  19J0870  19L0219  19L0478  19B0852  19D0226  19B0852R  19E0638  19F0229  19J0676  19K0946  19L0220  19L0490  19B0853  19B0853R  19F0232  19L0221  19L0492  19B0854  19D0227  19B0854R  19F0228  19C0638  19G0735  19L0222  19L0494  19D0228  19C0639  19G0738  19L0497  19D0229  19C0640  19G0739  19D0230  19C0641  19G0745  19C0642  19G0746  19L0498  19C0643  19G0747  19D0435  19G0524  19J0444  19K0947  19D0373  19G1066  19J0668  19K0948  19D0374  19G1067  19J0671  19K0949  19D0375  19G1068  19J0672  19K0950  19L0500  19D0433  19G1069  19J0673  19D0436  19G1070  19J0675  19K0951  19D0438  19D0437  19D0439  19D0376  19D0440  19D0441",
      "more_code_info": ""
    }
  ]
}