  "results": [
    {
      "status": "Terminated",
      "city": "Bothell",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84329",
      "recalling_firm": "Verathon, Inc.",
      "address_1": "20001 N Creek Pkwy",
      "address_2": "",
      "postal_code": "98011-8218",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: RI, IL, FL, SD, NY, MD, CA, OH, LA, AZ, GA, WA, OK, NC, AL, PA, TN, NJ, NV, MI, OR, TX, NM, VA, UT, NE, WI, CO, AR, MN, CT, ND, and KS.    OUS:  Canada",
      "recall_number": "Z-1471-2020",
      "product_description": "GlideScope Video Laryngoscope, Operations & Maintenance Manual, 0900-4940 REV 03",
      "product_quantity": "239 manuals",
      "reason_for_recall": "A typographical error in the Operations & Maintenance Manual incorrectly identifies the use of low level disinfection for the reusable video laryngoscope, specifically for the Titanium Blades.",
      "recall_initiation_date": "20190331",
      "center_classification_date": "20200311",
      "termination_date": "20211214",
      "report_date": "20200318",
      "code_info": "GlideScope Video Laryngoscope Operation and Maintenance Manual Catalog #:   0900-4940  Product can be further identified as In-Service USB 0901-0197    Affected Revision   Rev. 00 distributed January 1, 2019   February 28, 2019    GlideScope Video Laryngoscope Product Serial Numbers are included in Appendix A: Domestic and Foreign Consignee List (in Associated Documents)",
      "more_code_info": ""
    }
  ]
}