  "results": [
    {
      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72992",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed only in Puerto Rico.",
      "recall_number": "Z-1471-2016",
      "product_description": "Catalog Number:  900474  LAPAROSCOPY SURGICAL PACK  Used by medical staff to perform surgical procedures and  patient care procedures such as wound healing, suture removals, irrigate wounds, etc.",
      "product_quantity": "120",
      "reason_for_recall": "Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.",
      "recall_initiation_date": "20160107",
      "center_classification_date": "20160414",
      "termination_date": "20170503",
      "report_date": "20160420",
      "code_info": "LOTS:  SL15169 (150619284) 30-APR-16  SL15173 (150619372) 30-JUN-16  SL15195-02 (150619554) 30-JUN-16  SL15195-02 (150719804) 30-APR-16  SL15264 (150820330) 31-AUG-16  SL15274 (150920828) 31-AUG-16  SL15275 (150920828) 31-AUG-16",
      "more_code_info": ""
    }
  ]
}