      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67763",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of:  Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.",
      "recall_number": "Z-1464-2014",
      "product_description": "Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.",
      "product_quantity": "116 units",
      "reason_for_recall": "A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.",
      "recall_initiation_date": "20140306",
      "center_classification_date": "20140416",
      "termination_date": "20141105",
      "report_date": "20140423",
      "code_info": "System code: 728323; Serial #: 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320031, 320032, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320074, 320075, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320085, 320086, 320087, 320088, 320089, 320090, 320092, 320093, 320095, 320096, 320097, 320098, 320099, 320101, 320102, 320103, 320104, 320105,320106, 320107, 320109, 320110, 320111, 320112, 320114, 320115, 320116, 320117, 320118, 320119, 320120, 320121, 320122, 336013 & 336015."
    }
  ]
}