  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bloomington",
      "address_1": "750 N Daniels Way",
      "reason_for_recall": "Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.",
      "address_2": "",
      "product_quantity": "2,267",
      "code_info": "All lots",
      "center_classification_date": "20180419",
      "distribution_pattern": "Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.",
      "state": "IN",
      "product_description": "Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275.    Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.",
      "report_date": "20180425",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cook Inc.",
      "recall_number": "Z-1461-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79713",
      "termination_date": "20200513",
      "more_code_info": "",
      "recall_initiation_date": "20180322",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}