{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Livermore",
      "address_1": "151 Lindbergh Ave Ste I",
      "reason_for_recall": "Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure of the TigerPaw System II fastener.",
      "address_2": "",
      "product_quantity": "~ 4,154 total units in US - both model numbers",
      "code_info": "Part number C-TP-1507 (7 connector)   All serial numbers/lots",
      "center_classification_date": "20150423",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of:Australia, Bahrain, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy Netherland, Norway, Poland, Qatar, Slovenia, Spain, Saudi Arabia, Sweden, Switzerland, Turkey and United Kingdom.",
      "state": "CA",
      "product_description": "TIGERPAW System II,   Part number C-TP-1507 (7 connector)    The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).",
      "report_date": "20150429",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Laax, Inc.",
      "recall_number": "Z-1461-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70853",
      "termination_date": "20170317",
      "more_code_info": "",
      "recall_initiation_date": "20150325",
      "postal_code": "94551-9497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}