  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hertford",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70754",
      "recalling_firm": "Del Mar Reynolds Medical, Ltd.",
      "address_1": "1 2 Harforde Court John Tate Road",
      "address_2": "Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US including the states of Louisiana, North Carolina, and Washington and the countries of Argentina, Brazil, China, Colombia, Ecuador, Finland, Great Britain, India, Libya, Mexico, Nicaragua, Oman, Philippines, Saudi Arabia, Slovakia, Sri Lanka, United Arab Emirates, and Vietnam.",
      "recall_number": "Z-1458-2015",
      "product_description": "Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators.     Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia.",
      "product_quantity": "1040 units total (398 in the US and 642 international)",
      "reason_for_recall": "Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.",
      "recall_initiation_date": "20150305",
      "center_classification_date": "20150420",
      "termination_date": "20151116",
      "report_date": "20150429",
      "code_info": "PN: 376-0561-00.",
      "more_code_info": ""
    }
  ]
}