  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90500",
      "recalling_firm": "NuVasive Inc",
      "address_1": "7475 Lusk Blvd",
      "address_2": "",
      "postal_code": "92121-5707",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: CA, WA, TX, IN, MO, MI, FL, VA, OR.  OUS: DE, NZ, SG, UK",
      "recall_number": "Z-1453-2022",
      "product_description": "Description/Part:   Simplify Disc Size 1, Ht 5/SM-5;  Simplify Disc Size 1, Ht 6/SM-6;  Simplify Disc Size 2, Ht 4/MD-4;  Simplify Disc Size 2, Ht 5/MD-5;  Simplify Disc Size 2, Ht 5, 5/MD-5L;  Simplify Disc Size 2, Ht 6/MD-6;  Simplify Disc Size 2, Ht 6, 5/MD-6L;  Simplify Disc Size 3, Ht 5/LG-5;  Simplify Disc Size 3, Ht 5, 5/LG-5L;  Simplify Disc Size 3, Ht 6/LG-6;  Simplify Disc Size 3, Ht 6, 5/LG-6L",
      "product_quantity": "1,963",
      "reason_for_recall": "Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.",
      "recall_initiation_date": "20220620",
      "center_classification_date": "20220722",
      "report_date": "20220803",
      "code_info": "Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132, W81133, W81633, W81625, W81611, W81828, W81971, W82544, W82543, W81034, W81975, W81580, W81612, W81621, W81131, W81613, W80909, W81623, W81634, W79466, W81033, W81134, W81600, W79259; MD-5/843285100061/W81136, W81135, W80704, W80700; MD-5L/843285100078/W80707, W80706; MD-6/843285100085/W80911, W81642, W81643, W82545, W82546; MD-6L/843285100092/W81822; LG-5/843285100108/W81138, W81141, W79465, W81994, W81142, W81626, W81143, W79257; LG-5L/843285100115/W79258, W80146, W80915, W82548; LG-6/843285100122/W82547; LG-6L/843285100139/W81827",
      "more_code_info": ""
    }
  ]
}