  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90487",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states: KS and ME.",
      "recall_number": "Z-1452-2022",
      "product_description": "SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090.  Used for counting, displaying, recording and locating the number of RFID-tagged absorbent articles used during surgical procedures.",
      "product_quantity": "44 units",
      "reason_for_recall": "When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.",
      "recall_initiation_date": "20220615",
      "center_classification_date": "20220721",
      "termination_date": "20240405",
      "report_date": "20220727",
      "code_info": "UDI-DI (GTIN): 07613327543094;  Serial Numbers: GK404E00828  GK404E00933  GK404E00959  GK404E01157  GK405E00282  GK405E00642  GK405E00671  GK405E00675  GK405E00693  GK405E00707  GK405E00757  GK405E00827  GK405E00857  GK405E00914  GK403E00537  GK404E00494  GK405E00176  GK405E00655  GK405E00723  GK403E00031  GK404E00638  GK404E00640  GK404E00700  GK404E00866  GK404E00887  GK404E00945  GK404E01113  GK404E01119  GK404E01121  GK404E01123  GK404E01170  GK404E01171  GK405E00602  GK49CE00058  GK404E00032  GK404E00034  GK404E00043  GK404E00262  GK404E00287  GK404E00287  211071  211115  211265"
    }
  ]
}