  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67762",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution including  San Juan PR",
      "recall_number": "Z-1449-2014",
      "product_description": "Cobas c8000 Modular Analyzer Series Cobas c502,  Part Number 5964067001,  a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.",
      "product_quantity": "76 units",
      "reason_for_recall": "The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c pack MULTI or empty pre-labeled cobas c packs. The ICVC feature allows the reagent probe to dive into the reagent cassette and confirm the reagent level in the bottles. This ensures",
      "recall_initiation_date": "20140313",
      "center_classification_date": "20140410",
      "termination_date": "20180110",
      "report_date": "20140416",
      "code_info": "not applicable"
    }
  ]
}