{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "Potential risk for helium gas inside the MR examination room during a magnet quench",
      "address_2": "",
      "product_quantity": "8,205 units in total",
      "code_info": "(Added 6/25/18): 561  548  719  734  698  681  541  830  563  717  852  633  574  929  741  942  572  697  649  544  757  647  725  888  516  818  682  537  747  718  751  726  769  502  625  514  761  778  577  554  848  560  624  527  525  523  773  670  532  620  593  558  793  616  505  758  636  672  823  947  590  540  724  653  801  568  646  855  576  545  580  581  506  693  559  551  925  612  623  530  777  917  907  831  652  613  632  675  849  604  944  820  920  650  539  654  941  928  713  657",
      "center_classification_date": "20180418",
      "distribution_pattern": "Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia,   Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile,  China, Colombia, Congo (Democratic Republic of the), Costa Rica,   C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic,  Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary,  Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan,  Kazakhstan, Kenya, Republic of Korea - South Korea, Kuwait, Kyrgyzstan,   Latvia, Lebanon, Lesotho, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mauritius, Mexico, Monaco, Mongolia,  Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Caledonia,   New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of  Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Qatar,   Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore,  Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland,  Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Venezuela,  Viet Nam, Yemen & Zimbabwe.",
      "state": "MA",
      "product_description": "T5 Model 78104     Product Usage:  Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.",
      "report_date": "20180425",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-1442-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79451",
      "termination_date": "20200410",
      "more_code_info": "",
      "recall_initiation_date": "20180316",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}