      "status": "Terminated",
      "city": "Lake Oswego",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87549",
      "recalling_firm": "BIOTRONIK Inc",
      "address_1": "6024 Jean Rd",
      "address_2": "N/A",
      "postal_code": "97035-5571",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide  OUS: Worldwide",
      "recall_number": "Z-1441-2021",
      "product_description": "BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.",
      "product_quantity": "N/A",
      "reason_for_recall": "There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)",
      "recall_initiation_date": "20210308",
      "center_classification_date": "20210421",
      "termination_date": "20230425",
      "report_date": "20210428",
      "code_info": "serial numbers: 60705317  60707977  60708912  60709157  60709393  60709408  60709992  60710002  60710004  60710006  60710009  60710010  60710202  60710203  60711081  60711093  60711108  60711123  60711124  60711130  60711285  60711293  60711297  60711301  60711624  60711628  60711633  60711641  60711653  60711655  60711656  60711658  60711659  60711660  60711661  60712136  60712137  60712142  60712164  60712165  60712171  60712212  60712362  60712364  60712366  60712381  60712433  60712434  60712440  60712453  60712454  60712455  60712456  60712632  60712635  60712662  60712667  60712668  60712677  60712678  60712679  60712680  60712681  60712683  60712688  60712690  60713619  60713623  60713624  60713625  60713626  60713628  60713629  60713630  60713631  60713877  60713878  60713879  60713881  60713882  60713883  60713887  60713891  60713894  60713899  60713906  60713907  60713908  60713910  60713911  60713913  60713917  60713922  60713928  60713929  60713935  60713939  60713940  60713941  60713945  60713947  60713949  60713956  60713957  60713958  60713960  60713961  60713962  60713964  60713965  60713966  60713968  60713969  60713971  60713975  60713976  60713977  60713978  60713983  60713984  60713990  60713991  60713992  60713993  60713994  60713995  60714192  60714193  60714195  60714196  60714198  60714200  60714201  60714202  60714203  60714204  60714206  60714209  60714210  60714212  60714215  60714217  60714221  60714223  60714226  60714227  60714232  60714233  60714235  60714237  60714240  60714241  60714244  60714245  60714249  60714250  60714586  60714587  60714592  60714594  60714599  60714601  60714602  60714605  60714608  60714610  60714772  60714773  60714774  60714776  60714777  60714787  60714790  60714791  60714793  60714796  60714801  60714893  60714894  60714896  60714897  60714899  60714900  60714902  60714903  60714905  60714907  60714908  60714911  60714915  60714919  60714924  60714927  60714929  60714931  60714936  60714941  60714942  60714946  60714948  60714951  60714957  60714964  60714965  60714966  60714969  60714970  60715193  60715202  60715204  60715205  60715214  60715215  60715216  60715217  60715218  60715219  60715220  60715221  60715223  60715225  60715226  60715229  60715230  60715231  60715233  60715234  60715236  60715240  60715241  60715246  60715247  60715248  60715249  60715251  60715256  60715257  60715482  60715490  60715502  60715505  60715506  60715507  60715508  60715524  60715702  60715703  60715706  60715708  60715709  60715710  60715715  60715718  60715719  60715720  60715721  60715731  60715732  60715736  60715740  60715745  60715750  60715763  60715768  60715769  60715964  60715968  60715969  60715970  60715973  60715976  60715977  60715979  60715980  60715981  60715994  60715996  60715997  60716000  60716001  60716003  60716004  60716006  60716008  60716009  60716013  60716019  60716412  60716415  60716416  60716428  60716431  60716432  60716436  60716443  60716447  60716451  60716458  60716464  60716477  60716480  60716486  60716490  60716496  60716500  60716503  60716504  60716505  60716508  60716516  60716519  60716521  60716522  60716529  60716937  60716938  60716939  60716942  60716944  60716945  60716946  60716950  60716951  60716954  60716962  60716963  60716966  60716969  60716971  60716974  60716976  60716978  60716979  60716982  60716983  60716984  60716985  60716986  60716987  60716988  60716990  60716991  60716993  60716998  60717557  60717559  60717560  60717561  60717563  60717564  60717567  60717568  60717569  60717570  60717573  60717575  60717576  60717582  60717583  60717584  60717585  60717586  60717590  60717592  60717593  60717595  60717596  60717597  60717599  60717600  60717603  60717605  60717608  60717624  60717627  60717628  60717630  60717841  60717842  60717843  60717844  60717845  60717847  60717849  60717859  60717861  60717867  60717872  60717881  60717887  60717888  60717894  60717896  60717898  60717907  60717910  60718133  60718135  60718137  60718138  60718139  60718372  60718374  60718375  60718376  60718377  60718379  60719917  60719919  60719920  60719922  60719925  60719929  60719932  60719933  60719934  60719935  60719942  60719944  60719945  60719946  60719952  60719953  60719954  60719955  60719956  60719959  60719961  60719964  60719965  60719968  60719971  60719974  60720137  60720138  60720139  60720141  60720142  60720144  60720145  60720148  60720150  60720151  60720154  60720159  60720160  60720165  60720166  60720260  60720264  60720265  60720266  60720267  60720269  60720270  60720271  60720272  60720273  60720276  60720277  60720278  60720279  60720280  60720282  60720283  60720285  60720287  60720289  60720291  60720292  60720293  60720294  60720295  60720296  60720300  60720303  60720305  60720308  60720309  60720312  60720319  60720320  60720321  60720322  60720324  60720325  60720326  60721227  60721228  60721236  60721245  60721246  60721247  60721248  60721256  60721271  60721328  60721330  60721332  60721334  60721335  60721338  60721348  60721350  60721352  60721353  60721354  60721355  60721357  60721358  60721360  60721361  60721362  60721363  60721364  60721366  60721367  60721374  60721375  60721378  60721379  60721381  60721384  60721387  60721389  60721390  60721398  60721401  60721403  60721404  60721405  60721406  60722237  60722238  60722240  60722244  60722247  60722258  60722261  60722262  60722264  60722268  60722273  60722274  60722277  60722279  60722280  60722285  60722288  60722291  60722295  60722788  60722793  60722795  60722797  60722798  60722802  60722803  60722806  60722967  60722968  60722972  60722975  60722978  60722980  60722983  60722984  60722997  60722999  60723000  60723003  60723004  60723006  60723007  60723009  60723012  60723013  60723018  60723019  60723020  60723023  60723024  60723025  60723033  60723034  60723035  60723308  60723316  60723318  60723321  60723327  60723332  60723333  60724590  60724593  60724595  60724597  60724601  60724602  60724604  60724605  60724607  60724609  60724610  60724611  60724618  60724626  60724627  60724628  60724629  60724634  60724636  60724637  60724638  60724639  60724645  60724650  60724651  60724653  60724666  60724769  60724770  60724772  60724775  60724783  60724785  60724789  60724790  60724792  60724793  60724795  60724796  60724803  60724805  60724807  60724808  60724811  60724814  60724816  60724822  60724823  60724824  60724826  60724830  60724831  60724834  60724899  60724902  60724904  60724907  60724908  60724911  60724913  60724917  60724919  60724923  60724933  60724934  60724936  60724939  60724946  60724948  60725240  60725241  60725404  60725410  60725415  60725418  60725419  60725421  60725422  60725426  60725427  60725430  60725431  60725436  60725439  60725444  60725446  60725447  60725448  60725449  60725454  60725455  60725459  60725460  60725462  60725465  60725470  60725474  60725476  60725479  60725481  60725485  60725486  60725487  60725490  60725491  60725493  60725550  60725551  60725555  60725556  60725561  60725562  60725563  60725566  60725578  60725579  60725582  60725583  60725587  60725588  60725591  60725592  60725594  60725596  60725598  60725599  60725602  60725606  60725608  60725610  60725611  60725612  60725615  60725616  60725618  60725619  60725620  60725624  60725625  60725631  60725632  60725633  60725727  60725730  60725731  60725733  60725735  60725737  60725739  60725740  60725741  60725742  60725745  60725749  60725757  60725758  60725763  60725811  60725812  60725813  60725814  60726497  60726500  60726504  60726507  60726508  60726509  60726510  60726513  60726516  60726519  60726529  60726539  60726540  60726543  60726544  60726545  60726550  60726564  60726565  60726851  60726853  60726855  60726858  60726862  60726867  60726869  60726870  60726873  60726874  60726875  60726876  60726878  60726880  60726887  60726895  60726898  60726899  60726903  60726913  60727160  60727163  60727164  60727166  60727172  60727173  60727174  60727175  60727179  60727185  60727200  60727210  60727211  60727213  60727214  60727221  60727223  60727224  60727226  60727229  60727233  60727241  60727243  60727247  60727249  60727385  60727388  60727389  60727393  60727395  60727396  60727398  60727400  60727403  60727404  60727405  60727406  60727407  60727412  60727414  60727418  60727420  60727424  60727425  60727426  60727429  60727444  60727446  60727448  60727449  60727451  60727795  60727799  60727800  60727801  60727803  60728039  60728043  60728044  60728046  60728048  60728049  60728051  60728053  60728055  60728057  60728819  60728820  60728831  60728832  60728834  60728836  60728851  60728853  60728867  60728871  60728885  60728889  60728895  60728897  60728916  60728937  60728990  60728999  60729006  60729007  60729017  60729042  60729045  60729062  60729064  60729071  60729072  60729075  60729080  60729124  60729125  60729126  60729128  60729809  60729810  60729812  60729814  60729815  60729817  60730300  60730302  60730304  60730305  60730324  60730342  60730345  60730346  60730353  60730355  60730365  60730367  60730368  60730590  60730592  60730595  60730597  60730598  60730603  60730606  60730610  60730618  60730625  60730648  60730649  60730652  60730661  60730665  60730666  60731369  60731375  60731379  60731391  60731399  60731407  60731409  60731411  60731412  60731413  60731427  60731434  60731436  60731438  60732470  60732472  60732473  60732476  60732479  60732492  60732497  60732505  60732506  60732511  60732514  60732518  60732521  60732525  60732531  60732533  60732536  60733326  60733329  60733342  60733344  60733353  60733354  60733356  60733360  60733363  60733364  60733365  60733375  60733494  60733511  60733522  60733536  60733542  60733661  60733663  60733664  60733673  60733677  60733850  60733856  60733864  60733865  60733868  60733869  60733879  60733881  60733886  60733891  60733893  60733896  60733899  60733902  60733906  60734474  60734479  60734485  60734486  60734531  60734540  60734555  60734556  60734558  60734564  60734573  60734651  60734658  60734659  60734661  60734662  60734664  60734665  60734668  60734670  60734673  60734681  60734687  60734689  60734692  60734694  60734695  60734696  60734700  60734702  60734710  60734717  60734720  60734721  60734722  60734723  60734727  60734728  60734732  60734735  60734737  60734738  60734739  60735326  60735333  60735337  60735342  60735343  60735348  60735349  60735351  60735352  60735353  60735362  60735363  60735364  60735368  60735369  60735374  60735376  60735378  60735380  60735382  60735384  60735385  60735387  60735388  60735835  60735943  60735945  60735947  60735961  60735969  60735971  60735973  60735981  60735990  60735992  60735993  60736195  60736203  60736204  60736206  60736211  60736212  60736217  60736218  60736220  60736223  60736224  60736229  60736241  60736247  60736249  60736507  60736511  60736512  60736517  60736518  60736519  60736527  60736531  60736534  60736540  60736544  60737590  60737592  60737593  60737604  60737612  60737613  60737614  60737624  60737632  60737635  60737685  60737686  60737694  60737696  60737697  60737699  60737702  60737704  60737706  60737709  60737719  60737724  60738001  60738003  60738006  60738007  60738014  60738017  60738018  60738021  60738026  60738031  60738033  60738035  60738036  60738040  60738046  60738231  60738235  60738237  60738241  60738242  60738243  60738246  60738251  60738252  60738254  60738257  60738262  60738263  60738267  60738268  60738269  60738270  60738273  60738275  60738276  60738278  60738279  60738280  60738281  60738282  60738284  60738437  60738439  60738440  60738450  60738452  60738454  60738461  60738462  60738463  60738466  60738712  60738717  60738725  60738812  60738815  60738816  60738820  60738821  60738822  60738825  60738826  60738828  60738829  60738832  60738835  60738838  60738938  60738947  60738948  60738949  60738952  60738953  60738954  60738955  60738962  60738964  60738965  60738966  60738968  60738970  60738971  60738972  60738978  60738979  60738980  60738981  60738983  60738984  60739407  60739413  60739415  60739416  60739421  60739423  60739424  60739427  60739431  60739433  60739434  60739435  60739438  60739442  60739443  60739444  60739447  60739450  60739498  60739500  60739504  60739823  60739827  60739831  60739839  60739843  60739848  60739850  60739854  60739856  60739857  60739859  60740157  60740159  60740161  60740165  60740166  60740172  60740174  60740177  60740178  60740179  60740180  60740181  60740218  60740226  60740227  60740233  60740237  60740255  60740260  60740270  60740273  60740284  60740288  60740291  60740296  60740301  60740303  60740304  60740310  60740475  60740476  60740477  60740482  60740485  60740487  60740488  60740491  60740492  60740494  60740496  60740498  60740499  60740517  60740518  60740529  60740530  60740531  60740532  60740546  60740550  60740577  60740589  60740596  60740607  60740609  60740610  60740611  60740612  60740614  60740616  60740624  60740695  60740699  60740709  60740713  60740714  60740722  60740729  60740733  60740734  60740967  60740968  60740971  60740984  60740986  60740992  60741009  60741015  60741017  60741019  60741022  60741024  60741030  60741033  60741037  60741045  60741050  60741055  60741058  60741061  60741269  60741270  60741277  60741278  60741285  60741287  60741290  60741291  60741292  60741296  60741304  60741309  60741311  60741316  60741321  60741324  60741327  60741341  60741344  60741620  60741624  60741627  60741632  60741638  60741643  60741646  60741735  60741756  60741760  60741763  60741764  60741765  60741769  60741773  60741775  60741778  60741780  60741795  60741796  60741806  60741807  60741814  60741820  60741822  60744300  60744932  60745802  60745803  60745812  60746817  60746818  60746820  60746855  60746866  60746867  60746868  60746869  60746870  60746871  60746872  60746873  60747032  60747034  60747035  60747036  60747038  60747039  60747040  60747042  60747050  60747051  60747053  60747055  60747062  60747065  60747066  60747067  60747068  60747069  60747070  60747074  60747077  60747079  60747091  60747094  60747095  60747096  60747098  60747099  60747100  60747104  60747108  60747395  60747397  60747401  60747404  60747413  60747416  60747419  60747420  60747421  60747422  60747423  60747426  60747427  60747428  60747440  60747444  60747446  60747452  60747453  60747455  60747459  60747648  60747650  60747651  60747655  60747656  60747657  60747658  60747660  60747661  60747665  60747668  60747670  60747672  60747673  60747674  60747777  60747786  60747789  60747790  60747799  60747800  60747802  60747836  60747837  60747840  60747854  60747856  60747857  60747861  60747862  60747879  60747881  60747882  60747884  60748035  60748040  60748042  60748044  60748046  60748049  60748050  60748053  60748055  60748057  60748059  60748061  60748068  60748070  60748073  60748076  60748077  60748081  60748094  60748095  60748355  60748356  60748357  60748360  60748361  60748362  60748364  60748365  60748367  60748368  60748369  60748401  60748402  60748403  60748404  60748407  60748408  60748410  60748411  60748412  60748415  60748416  60748417  60748418  60748419  60748421  60748422  60748423  60748424  60748425  60748426  60748427  60748428  60748430  60748434  60748438  60748441  60748443  60750116  60750118  60750128  60750135  60750149  60750155  60750165  60750172  60750175  60750180  60750181  60750183  60750184  60750193  60750194  60750196  60750210  60750212  60750218  60750227  60750229  60750241  60750251  60750255  60750257  60750260  60750556  60750558  60750560  60750568  60750570  60750572  60750573  60750576  60750588  60750589  60750591  60750592  60750603  60750605  60750786  60750787  60750789  60750790  60750792  60750793  60750794  60750795  60750796  60750798  60750799  60750805  60750807  60750810  60750814  60750821  60750825  60750828  60750829  60750830  60750837  60750838  60750839  60750840  60750842  60750843  60750847  60750849  60750852  60751017  60751020  60751023  60751029  60751032  60751034  60751037  60751039  60751041  60751042  60751145  60751147  60751148  60751153  60751156  60751157  60751160  60751162  60751163  60751164  60752801  60752802  60752803  60752804  60752808  60752809  60752810  60752811  60752823  60752825  60752829  60752832  60752834  60752835  60752837  60752839  60753046  60753047  60753048  60753049  60753053  60753054  60753056  60753059  60753061  60753062  60753063  60753064  60753070  60753074  60753076  60753081  60753083  60753085  60753086  60753087  60753092  60753096  60753099  60753100  60753101  60753105  60753106  60753108  60753110  60753996  60753998  60754000  60754004  60754005  60754007  60754008  60754010  60754011  60754013  60754020  60754024  60754026  60754030  60754031  60754332  60754337  60755021  60755044  60755055  60755062  60755067  60755089  60755090  60755271  60755272  60755274  60755275  60755771  60755774  60755777  60755780  60755782  60755785  60755786  60755788  60755789  60755791  60755794  60755798  60755803  60755805  60755807  60755808  60755811  60755818  60755822  60755826  60756030  60756043  60756046  60756048  60756050  60756052  60756053  60756058  60756064  60756065  60756084  60756086  60756087  60756088  60756096  60756099  60756100  60756101  60756102  60756106  60756108  60756145  60756149  60756150  60756154  60756157  60756158  60756167  60756349  60756356  60756361  60756363  60756364  60756370  60756371  60756372  60756376  60756378  60756381  60756387  60756648  60756651  60756656  60756975  60756976  60756983  60757121  60757122  60757123  60757129  60757147  60757148  60757152  60757153  60757154  60757156  60757157  60757158  60757161  60757162  60757168  60757176  60757179  60757180  60757659  60757664  60757665  60757667  60757679  60757687  60757689  60757690  60757693  60757694  60757695  60757828  60757832  60757835  60757836  60758259  60758264  60758268  60758277  60758281  60758283  60758286  60758295  60758448  60758449  60758451  60758452  60758454  60758455  60758459  60758461  60758463  60758465  60758548  60758549  60758551  60758554  60758557  60758558  60758561  60758562  60758564  60758567  60758569  60758570  60758572  60758578  60758584  60758594  60758599  60758604  60759767  60759780  60759784  60760050  60760053  60760057  60760063  60760583  60760802  60760814  60760837  60760838  60760839  60760854  60761026  60761027  60761029  60761033  60761034  60761056  60761058  60761061  60761178  60761202  60761203  60761209  60761223  60761224  60761310  60761320  60761353  60761354  60761867  60761890  60761907  60762746  60762749  60762751  60762752  60762778  60762927  60762928  60762929  60762930  60762933  60762936  60762937  60762938  60762939  60762942  60762944  60762946  60762947  60762948  60762949  60762950  60762951  60762953  60762966  60762969  60762970  60762974  60762975  60762978  60762984  60763107  60763110  60763111  60763113  60763115  60763557  60763558  60763561  60763816  60763818  60763819  60763820  60763821  60763822  60763824  60763833  60763834  60763836  60763837  60763838  60763839  60763841  60763842  60763843  60763844  60763845  60763848  60763849  60763850  60763853  60763917  60763920  60763921  60763922  60763923  60763924  60763925  60763926  60763927  60763929  60763931  60763932  60763933  60763935  60763938  60763940  60763941  60763942  60763943  60763944  60763946  60763948  60763950  60763951  60763952  60763954  60764358  60764360  60764363  60764364  60765013  60765016  60765304  60765317  60765368  60765370  60765372  60765393  60765400  60765633  60765639  60765641  60765642  60765650  60765652  60765656  60766306  60766308  60766317  60766458  60766459  60766462  60766466  60766467  60766469  60766473  60766476  60766481  60766483  60766488  60766491  60766492  60766495  60766497  60766500  60766501  60766505  60766508  60766515  60766516  60766517  60767423  60767425  60767432  60767436  60767445  60767460  60767461  60767484  60767993  60767994  60768018  60768025  60768028  60768029  60768035  60768041  60768051  60768053  60768056  60768058  60768060  60768061  60768064  60768065  60768066  60768067  60768149  60768155  60768164  60768165  60768180  60768181  60768184  60768187  60768188  60768191  60768206  60768207  60768213  60768216  60768223  60768224  60768229  60768231  60768234  60768235  60768298  60768303  60768317  60768318  60768320  60768322  60768323  60768328  60768335  60768336  60768438  60768440  60768446  60768448  60768458  60768470  60768471  60768474  60768479  60768486  60768645  60768654  60768676  60768681  60768683  60768685  60768689  60768691  60768693  60768696  60768719  60768722  60768727  60768734  60768735  60768737  60768751  60768754  60768873  60768878  60768882  60768887  60768892  60768893  60768899  60768908  60768909  60769162  60769165  60769166  60769169  60769170  60769172  60769173  60769180  60769181  60769188  60769195  60769196  60769199  60769210  60769212  60769213  60769228  60769229  60769230  60769231  60769232  60769234  60769239  60769271  60769272  60769273  60769600  60769605  60769614  60769615  60769616  60769905  60769906  60769907  60769914  60769924  60769940  60769943  60769946  60769948  60769957  60769962  60769963  60769971  60769973  60769976  60770176  60770189  60770195  60770212  60770313  60770319  60770320  60770327  60770337  60770338  60770340  60770344  60770346  60770348  60770349  60770354  60770357  60770361  60770477  60770480  60770482  60770488  60770489  60770493  60770497  60770554  60770557  60770761  60770767  60770776  60770777  60770781  60770798  60770802  60770804  60770807  60770901  60770902  60770906  60770911  60770917  60770919  60770927  60770934  60770935  60771120  60771121  60771140  60771141  60771268  60771285  60771733  60771753  60771759  60771767  60771854  60771858  60771860  60771890  60771891  60772476  60772765  60772766  60772767  60773185  60773211  60773224  60773234  60773235  60773236  60773238  60773241  60773244  60773245  60773257  60773261  60773265  60773269  60773288  60773289  60773301  60773484  60773486  60773487  60773489  60773497  60773887  60773896  60773897  60773916  60773921  60773922  60773924  60773927  60773928  60773937  60774136  60774137  60774143  60774146  60774150  60774156  60774157  60774160  60774165  60774168  60774185  60774197  60774202  60774235  60774240  60774255  60774257  60774259  60774260  60774261  60774262  60774264  60774266  60774269  60774270  60774273  60774296  60774302  60774631  60775224  60775226  60775604  60775605  60775609  60775614  60775615  60775616  60775619  60775621  60775623  60775625  60775626  60775627  60775640  60775648  60775650  60775652  60775809  60775819  60775826  60775846  60775848  60776013  60776030  60776031  60776032  60776064  60776669  60776821  60776832  60776842  60776848  60776852  60776875  60776876  60777095  60777115  60777126  60777244  60777257  60777262  60777562  60777563  60777564  60777566  60777567  60777573  60777576  60777582  60778242  60778260  60778267  60778269  60778276  60778308  60778312  60778317  60778442  60778639  60778647  60778649  60778658  60778665  60778668  60779047  60779074  60779724  60779737  60779738  60779740  60779741  60779769  60780134  60780620  60780784  60780821  60780823  60780832  60780967  60780970  60780972  60780987  60780996  60783823  60784237  60785440  60785584  60785592  60785596  60786734  60786872  60788773  60788776  60788777  60788864  60788866  60788869  60788870  60788955  60788961  60788962  60788974  60788977  60788992  60788996  60789010  60789012  60789017  60789021  60789024  60789026  60789028  60789029  60789030  60789040  60789052  60789374  60789375  60789381  60789385  60789394  60789405  60789406  60789411  60789413  60789414  60789418  60789420  60789423  60789426  60789434  60789437  60789442  60789483  60789492  60789496  60789516  60789519  60789520  60789522  60789527  60789530  60789532  60789541  60789558  60789559  60789560  60789564  60789569  60789571  60790053  60790056  60790057  60790058  60790060  60790061  60790062  60790064  60790083  60790087  60790090  60791288  60791306  60791310  60791339  60791739  60791749  60791750  60791769  60791772  60791775  60791776  60791788  60791795  60791812  60791946  60791949  60791950  60791953  60791970  60791992  60794070  60794314  60794344  60794347  60794359  60794449  60794492  60794493  60794495  60794503  60794505  60794514  60794515  60794521  60794523  60794526  60794780  60794785  60794786  60794790  60794800  60794805  60794813  60794816  60796045  60796047  60796048  60796049  60796060  60796068  60796073  60796074  60797160  60798046  60798050  60798052  60798055  60798083  60798094  60798100  60798119  60798308  60798314  60798665  60798668  60798671  60798674  60798999  60799000  60799006  60799008  60799010  60799017  60799022  60799055  60799059  60800890  60800901  60800902  60800918  60800921  60800927  60800929  60800932  60800979  60801228  60801281  60808870  60808871  60808874  60808876  60808968  60809147  60809152  60714800  60714912  60716472  60719930  60720290  60721346  60722791  60722801  60724928  60725732  60726864  60734657  60735355  60737708  60741816  60751015  60753091  60755023  60756089  60758296  60763919  60768174  60768334  60769975  60771768  60774142  60774676  60775651  60788184  60788872  60788984  60789554  60791314  60791347  60798475  60798677",