  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67677",
      "recalling_firm": "Draeger Medical Systems, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution in the states of WI and NY.",
      "recall_number": "Z-1440-2014",
      "product_description": "Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833",
      "product_quantity": "3",
      "reason_for_recall": "The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.",
      "recall_initiation_date": "20140305",
      "center_classification_date": "20140410",
      "termination_date": "20141031",
      "report_date": "20140416",
      "code_info": "Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003,   ASEK-0004, ASEK-0005"
    }
  ]
}