{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dublin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65109",
      "recalling_firm": "Carl Zeiss Meditec, Inc.",
      "address_1": "5160 Hacienda Dr",
      "address_2": "N/A",
      "postal_code": "94568-7562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution.",
      "recall_number": "Z-1440-2013",
      "product_description": "Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C),      Product Usage:  A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.",
      "product_quantity": "6 units",
      "reason_for_recall": "Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.",
      "recall_initiation_date": "20130418",
      "center_classification_date": "20130531",
      "termination_date": "20130626",
      "report_date": "20130612",
      "code_info": "Serial Numbers:  6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813"
    }
  ]
}