  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mettawa",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84337",
      "recalling_firm": "Vyaire Medical",
      "address_1": "26125 N Riverwoods Blvd",
      "address_2": "N/A",
      "postal_code": "60045-3420",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.",
      "recall_number": "Z-1439-2020",
      "product_description": "bellavista 1000 ventilator, Catalog number 301.100.030.  bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system.  bellavista uses room air and high-pressure oxygen.  Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower.  Oxygen enters through a high pressure inlet.  An electronic mixer valve provides for the operator-set concentration.  Gas is supplied to the patient via the microprocessor controlled inspiratory valve.",
      "product_quantity": "116",
      "reason_for_recall": "The G6 bellavista 1000 US ventilators may experience intermittent failures:  Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.",
      "recall_initiation_date": "20191125",
      "center_classification_date": "20200304",
      "termination_date": "20230808",
      "report_date": "20200311",
      "code_info": "Serial numbers:  MB170104, MB170118, MB170151, MB170202, MB230000, MB230020, MB230024, MB230027, MB230031, MB230033, MB230035, MB230036, MB230037, MB230038, MB230039, MB230040, MB230041, MB230046, MB230048, MB230049, MB230052, MB230053, MB230054, MB230055, MB230056, MB230057, MB230058, MB230059, MB230060, MB230061, MB230062, MB230063, MB230064, MB230065, MB230066, MB230067, MB230068, MB230069, MB230070, MB230071, MB230072, MB230073, MB230074, MB230075, MB230076, MB230077, MB230078, MB230079, MB230080, MB230081, MB230082, MB230083, MB230084, MB230085, MB230086, MB230087, MB230088, MB230089, MB230090, MB230091, MB230092, MB230093, MB230094, MB230095, MB230097, MB230098, MB230099, MB230100, MB230101, MB230102, MB230104, MB230190, MB230191, MB230194, MB230195, MB230196, MB230197, MB230201, MB230204, MB230205, MB230207, MB230209, MB230210, MB230217, MB230219, MB230220, MB230221, MB230222, MB230227, MB230228, MB230229, MB230230, MB230231, MB230232, MB230233, MB230243, MB230244, MB230246, MB230247, MB230249, MB230251, MB230253, MB230256, MB230258, MB230260, MB230261, MB230262, MB230263, MB230266, MB230270, MB230274, MB230275, MB230277, MB230294, MB230297, MB230298."
    }
  ]
}