  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76189",
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "address_1": "1265 Grey Fox Rd",
      "address_2": "N/A",
      "postal_code": "55112-6929",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Internationally to Finland",
      "recall_number": "Z-1439-2017",
      "product_description": "CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,",
      "product_quantity": "20",
      "reason_for_recall": "20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.",
      "recall_initiation_date": "20161031",
      "center_classification_date": "20170310",
      "termination_date": "20190715",
      "report_date": "20170322",
      "code_info": "1071155, 1076826, 1081258, 1081260, 1081275, 1082475, 1082511, 1082512, 1090347, 1090348, 1090349, 1091767, 1091768, 1091769, 1091770, 1091771, 1092395, 1100186, 1100188, 1100189."
    }
  ]
}