  "results": [
    {
      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72992",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed only in Puerto Rico.",
      "recall_number": "Z-1438-2016",
      "product_description": "Catalog Number:  9001780  OPHTHALMIC SURGICAL PACK   Used by medical staff to perform surgical procedures and  patient care procedures such as wound healing, suture removals, irrigate wounds, etc.",
      "product_quantity": "168",
      "reason_for_recall": "Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.",
      "recall_initiation_date": "20160107",
      "center_classification_date": "20160414",
      "termination_date": "20170503",
      "report_date": "20160420",
      "code_info": "LOTS:  SL15169 (150518594) 30-JUN-16  SL15170 (150518592) 30-JUN-16  SL15199 (150719720) 30-JUN-16  SL15199 (150719719) 30-JUN-16  SL15203 (150719720) 30-JUN-16  SL15230 (150819721) 30-JUN-16  SL15231 (150819722) 30-JUN-16  SL15231 (150820230) 30-JUN-16  SL15243 (150820231) 30-JUN-16  SL15243 (150820229) 30-JUN-16  SL15259 (150920791) 30-JUN-16  SL15259 (150920917) 30-JUN-16",
      "more_code_info": ""
    }
  ]
}