  "results": [
    {
      "status": "Terminated",
      "city": "Portland",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65120",
      "recalling_firm": "TZ Medical Inc.",
      "address_1": "17750 SW Upper Boones Ferry Rd Ste 150",
      "address_2": "N/A",
      "postal_code": "97224-7086",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide distribution - USA (nationwide) and Chile.",
      "recall_number": "Z-1438-2013",
      "product_description": "adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1.    Product Usage:  Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring.",
      "product_quantity": "US: 10,410 pieces; Outside US: 100 pieces",
      "reason_for_recall": "TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly.    If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor",
      "recall_initiation_date": "20130430",
      "center_classification_date": "20130530",
      "termination_date": "20130905",
      "report_date": "20130605",
      "code_info": "Model P-211-M1 Lot numbers:   Y081712-17; Y081712-18; Y081712-19; Y081712-20; Y112612-08; Y112612-09; Y112612-10; Y112612-11;  Y112612-12; Y112612-13; and Y020113-07.    Model P-214-M1 Lot numbers:   Y081712-04; Y112612-24; Y112612-23; and Y031913-10."
    }
  ]
}