{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76459",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-1435-2017",
      "product_description": "MEVATRON M 7400/7440(MEX) Accelerator, Linear, Medical  The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer",
      "product_quantity": "35 systems",
      "reason_for_recall": "Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.",
      "recall_initiation_date": "20170125",
      "center_classification_date": "20170309",
      "termination_date": "20171006",
      "report_date": "20170315",
      "code_info": "Device Model Number 8319758  Serial numbers: 3009  2958  2863  2972  2646  3079  2516  2962  3055  1452  2702  2125  3066  2689  2228  2564  2507  1904  3075  2375  3013  3068  2596  3025  3049  2965  2045  3084  2855  2906  2292  2384  2067  2818  3089"
    }
  ]
}