{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84929",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV.  There was government distribution and no military distribution.    Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Republic, Germany, France, Georgia, Greece, Hong Kong, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1431-2020",
      "product_description": "Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.",
      "product_quantity": "12,875 units",
      "reason_for_recall": "Lot numbers and products inadvertently not included in the scope of the previous recall.  Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.",
      "recall_initiation_date": "20200103",
      "center_classification_date": "20200303",
      "termination_date": "20220317",
      "report_date": "20200311",
      "code_info": "Lot numbers VTCT0268, VTCT0269, VTCT0270, VTCT0271, VTCU0340, and VTCU0341.     UDI numbers:  (01)00801741086281(17)200601(10)VTCT0268, (01)00801741086281(17)200601(10)VTCT0269, (01)00801741086281(17)200601(10)VTCT0270, (01)00801741086281(17)200601(10)VTCT0271, (01)00801741086281(17)200601(10)VTCU0340, and (01)00801741086281(17)200601(10)VTCU0341."
    }
  ]
}