{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65194",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.",
      "recall_number": "Z-1431-2013",
      "product_description": "8mm Precise Bipolar Forceps used in conjunction with the da Vinci  Surgical System, Model IS1200.    Multiple use electrosurgical endoscopic instruments.",
      "product_quantity": "697 units - all products",
      "reason_for_recall": "Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser,  preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.",
      "recall_initiation_date": "20130508",
      "center_classification_date": "20130529",
      "termination_date": "20140218",
      "report_date": "20130605",
      "code_info": "Model number: 400110-10;  Lot numbers:  M10120807  M10120817  M10120831  M10120910  M10121008  M10121012  M10121024  M10121115  M10121123  M10121130  M10121206  M10121210  M10130124  M10130131  M10130205  M10130304  M10130319  M10130321"
    }
  ]
}