  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65118",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA Nationwide and countries of: AU, CA, DE, ES, GB, IT, PT,  and JP.",
      "recall_number": "Z-1430-2013",
      "product_description": "AU2700 Clinical Chemistry Analyzer  AU5400 Clinical Chemistry Analyzer    Perform automated analysis of serum, urine, and other body fluids, including whole blood.",
      "product_quantity": "108 units",
      "reason_for_recall": "Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.",
      "recall_initiation_date": "20130426",
      "center_classification_date": "20130529",
      "termination_date": "20140804",
      "report_date": "20130605",
      "code_info": "All Serial Numbers    Model Numbers: AU2700, AU27000ISE, AU5421, AU5421ISE, AU5431.    Part Numbers:  E0750053  N3663400   N3663700   N3664000   N3664200   N3664400   N3664600   N3664800   N3878500   1465600   64270G   E61DAU2700  N3663500   N3663600   N3663800   N3663900   N3664100   N3664300   N3664500   N3664700   N3664900   N3878600   O9102700  N3665000   N3665100   N3665200   N3665300   N3665400   N3878300   64542   A94504   N1683100   AU5431-U1   N3665500   N3665600   N3665700   N3665800   N3665900   N3878400"
    }
  ]
}