  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84832",
      "recalling_firm": "Beckman Coulter Inc",
      "address_1": "2470 Faraday Ave",
      "address_2": "N/A",
      "postal_code": "92010-7224",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Alabama,  Alaska,  Arizona,  Arkansas,  California,  Colorado,  Connecticut,  Delaware,  Florida,  Georgia,  Hawaii,  Idaho,  Illinois,  Indiana,  Iowa,  Kansas,  Kentucky,  Louisiana,  Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming, Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas,Utah,Virginia,Washington,West Virginia, Wisconsin,Wyoming      Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Curacao, Czech Republic, Egypt, Eritrea, Finland,  France, Germany, Ghana, Greece, Guatemala, Guyana, Hong Kong, India, Ireland, Italy, Japan, Kuwait, Lebanon, Lithuania,  Malaysia, Mayotte, Mexico, Netherlands, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, San Marino, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam",
      "recall_number": "Z-1426-2020",
      "product_description": "Total Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300 tests/cartridge) and 476861 (400 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233075 (300 tests/cartridge) and 15099590625597 (400 tests/cartridge)",
      "product_quantity": "25,383 Reagents",
      "reason_for_recall": "It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.",
      "recall_initiation_date": "20190815",
      "center_classification_date": "20200302",
      "termination_date": "20240228",
      "report_date": "20200311",
      "code_info": "All Lot Numbers"
    }
  ]
}