  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lenexa",
      "state": "KS",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87495",
      "recalling_firm": "Remel Inc",
      "address_1": "12076 Santa Fe Trail Dr",
      "address_2": "N/A",
      "postal_code": "66215-3519",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1425-2021",
      "product_description": "Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503     Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative procedures by the Clinical and Laboratory Standards Institute (CLSI) as the growth medium for antimicrobial disk diffusion tests with Haemophilus species.",
      "product_quantity": "19 units",
      "reason_for_recall": "When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.",
      "recall_initiation_date": "20210310",
      "center_classification_date": "20210413",
      "report_date": "20210421",
      "code_info": "Lot Numbers: Lot 220711 Exp. 2021-2-17"
    }
  ]
}